Real-time nucleic acid sequence-based amplification (NASBA) using molecular beacon technology (NASBA-beacon) was compared to standard NASBA with postamplification hybridization using electrochemiluminescently labeled probes (NASBA-ECL) for detection of enteroviruses (EV) in 133 cerebrospinal fluid and 27 stool samples. NASBA-ECL and NASBA-beacon were similar in sensitivity, detecting 55 (100%) and 52 (94.5%) EV-positive samples, respectively. There were no false positives. Both NASBA assays were significantly more sensitive than culture. Real-time NASBA-beacon reagents and equipment rental were more expensive than those for NASBA-ECL; however, time to result was shortened by 1.5 h, hands-on time was reduced by 25 min, and the assay was much simpler for technologists to learn and perform.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC1169110PMC
http://dx.doi.org/10.1128/JCM.43.7.3136-3139.2005DOI Listing