Cyclotorsional eye movements during a simulated PRK procedure.

Eye (Lond)

Clinica Oculistica, Universita' di Torino, Dipartimento di Fisiopatologia Clinica, V Juvarra 19, Torino, Italy.

Published: July 2006

Purpose: Eye alignment is a serious concern when performing corneal surface ablation. Although several excimer lasers monitor horizontal and vertical movements, little is known about the potential impact of cyclotorsional movements. Dynamic cyclotorsions have been measured on 10 emmetropic subjects during a simulated PRK treatment.

Methods: Cyclotorsional eye movements were measured using a 3D videoculographic system (SMI) in 10 emmetropic subjects in upright and supine position. The subjects were wearing an eyelid speculum and were asked to fixate on a 1 degrees red spot. The fixation target was either in focus or optically degraded by electronic filters. Data were collected at the University Eye Clinic, Torino, Italy, and the different setting measurements were compared using Student's t-test for paired data.

Results: Cyclotorsions were significantly higher when subjects were fixating an optically degraded stimulus (upright: P=0.04; supine: P=0.0002). The cyclotorsional movements ranged from -13 to 17 degrees. A significant difference in eye orientation was present when changing from the upright to the supine position (P=0.03). Even when correcting for this positional error, significantly higher cyclotorsional movements were detected in the supine position (in focus: P=0.0043; optically degraded: P=0.0008). The torsional range was from -14.85 to 14.55 degrees.

Conclusions: The high range of cyclotorsional eye movements during a simulated PRK procedure suggests that 3D tracking could improve surface ablation results when treating high astigmatism or when wavefront guided.

Download full-text PDF

Source
http://dx.doi.org/10.1038/sj.eye.6701994DOI Listing

Publication Analysis

Top Keywords

cyclotorsional eye
12
eye movements
12
simulated prk
12
cyclotorsional movements
12
supine position
12
optically degraded
12
movements simulated
8
prk procedure
8
surface ablation
8
emmetropic subjects
8

Similar Publications

Idiopathic superior oblique muscle palsy is a major type of paralytic, non-comitant strabismus and presents vertical and cyclo-torsional deviation of one eye against the other eye, with a large vertical fusion range and abnormal head posture such as head tilt. Genetic background is considered to play a role in its development, as patients with idiopathic superior oblique muscle palsy have varying degrees of muscle hypoplasia and, rarely, the complete absence of the muscle, that is, aplasia. In this study, whole genome sequencing was performed, and single nucleotide variations and short insertions/deletions (SNVs/InDels) were annotated in two patients each in three small families (six patients in total) with idiopathic superior oblique muscle palsy, in addition to three normal individuals in one family.

View Article and Find Full Text PDF

Neurophysiology of the saccadic system: The reticular formation.

Prog Brain Res

March 2022

Late Professor of Ophthalmology, Biomedical Engineering and Neuroscience, Johns Hopkins University School of Medicine, Baltimore, MD, United States.

This chapter discusses the neurophysiology and function of subcortical circuits and cortical areas involved in saccade generation. While cells within the different nuclei of the brainstem reticular formation shape the temporal details of ipsiversive horizontal and vertical/cyclotorsional saccade components, the cerebellar flocculus, vermis and fastigial nucleus are thought to modulate these saccadic waveforms. Burst neurons in the deep layers of the superior colliculus encode the saccade vector in the contralateral field by a localized population in a motor-error map.

View Article and Find Full Text PDF

Clinical measurements of normative subjective cyclotorsion and cyclofusion in a healthy adult population.

Acta Ophthalmol

March 2020

Department of Clinical Neuroscience, Institute of Neuroscience and Physiology, Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden.

Purpose: To investigate normative subjective cyclotorsion values and cyclofusion ranges in a healthy adult population.

Methods: A cross-sectional investigation was performed in 120 healthy, non-strabismic adults, 60 men and 60 women in the age range of 18-69 years. All subjects were assessed for cyclotorsion using the synoptophore and the single Maddox rod (SMR) methods.

View Article and Find Full Text PDF

Purpose: To evaluate static and dynamic cyclotorsions during photorefractive keratotomy (PRK) surgery in refractive surgery candidates and their correlations with preoperative factors.

Methods: This cross-sectional case series was performed in 138 eyes of 77 patients who underwent PRK surgery by Technolas 217z100. Iris registration was used to evaluate the degree of static and dynamic cyclotorsion.

View Article and Find Full Text PDF

Cyclotorsion measurement using scleral blood vessels.

Comput Biol Med

August 2017

Hitit University, Faculty of Medicine, Department of Ophthalmology, 19030, Çorum, Turkey. Electronic address:

Background And Objectives: Measurements of the cyclotorsional movement of the eye are crucial in refractive surgery procedures. The planned surgery pattern may vary substantially during an operation because of the position and eye movements of the patient. Since these factors affect the outcome of an operation, eye registration methods are applied in order to compensate for errors.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!