The purpose of this study is to describe and evaluate a new analytical model for Varian enhanced dynamic wedge factors at off-center points. The new model was verified by comparing measured and calculated wedge factors for the standard set of wedge angles (i.e., 15 degrees, 30 degrees, 45 degrees and 60 degrees), different symmetric and asymmetric fields, and two different photon energies. The maximum difference between calculated and measured wedge factors is less than 2%. The average absolute difference is within 1%. The obtained results indicate that the suggested model can be useful for independent dose calculation with enhanced dynamic wedges.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.1872532DOI Listing

Publication Analysis

Top Keywords

wedge factors
12
degrees degrees
12
factors off-center
8
off-center points
8
enhanced dynamic
8
analytical representation
4
representation varian
4
varian edw
4
factors
4
edw factors
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!