Purpose: To compare the posterior capsule opacification (PCO) inhibiting effect of the sharp posterior optic edge design of the Sensar OptiEdge AR40e intraocular lens (IOL) with that of the double-round edge design of the Sensar AR40 IOL over a period of 3 years.

Setting: Department of Ophthalmology, Medical University of Vienna, Vienna, Austria.

Methods: Fifty-three patients with bilateral age-related cataract (106 eyes) were included in this randomized prospective bilateral patient- and examiner-masked clinical trial with intraindividual comparison. Each study patient had cataract surgery in both eyes and received a Sensar AR40 IOL (anterior and posterior round optic edges) in 1 eye and a Sensar OptiEdge AR40e IOL (round anterior and sharp posterior optic edge) in the other eye. Follow-up examinations were at 1 week, 1 and 6 months, and 1, 2, and 3 years. Digital retroillumination images were taken of each eye. The amount of PCO was assessed subjectively at the slitlamp and objectively using automated image-analysis software (AQUA) 1, 2, and 3 years after surgery.

Results: The sharp-edged AR40e lens showed significantly less regeneratory and fibrotic PCO 1, 2, and 3 years after surgery. The mean AQUA PCO score was 2.18 for the AR40 and 1.00 for the AR40e lens after 1 year; 2.94 and 1.56 after 2 years, and estimated at 3.13 and 1.77, respectively, after 3 years (scale 0 to 10; P<.001). The neodymium:YAG laser capsulotomy rate was significantly higher in the AR40 group. The AR40e lens also led to less peripheral fibrotic PCO. There was no significant difference in complaints concerning edge glare between the groups.

Conclusion: Compared with the AR40 IOL, the sharp posterior optic edge design of the Sensar OptiEdge AR40e IOL led to significantly less PCO 1, 2, and 3 years postoperatively. In contrast with most past studies on PCO-preventative factors, this study varied only 1 parameter, namely the optic edge design, and the main outcome measure (PCO) was assessed with an objective image-analysis system.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcrs.2004.09.053DOI Listing

Publication Analysis

Top Keywords

optic edge
12
edge design
12
intraocular lens
8
posterior capsule
8
capsule opacification
8
sharp posterior
8
posterior optic
8
design sensar
8
sensar optiedge
8
optiedge ar40e
8

Similar Publications

This study aims to determine the extent to which coating composition and workpiece properties impact machinability and tool selection when turning Compacted Graphite Iron (CGI) under extreme roughing conditions. Two CGI workpieces, differing in pearlite content and graphite nodularity, were machined at a cutting speed of 180 m/min, feed rate of 0.18 mm/rev, and depth of cut of 3 mm.

View Article and Find Full Text PDF

Intraoperative Augmented Reality for Vitreoretinal Surgery Using Edge Computing.

J Pers Med

January 2025

Department of Ophthalmology, Mayo Clinic, Rochester, MN 55905, USA.

: Augmented reality (AR) may allow vitreoretinal surgeons to leverage microscope-integrated digital imaging systems to analyze and highlight key retinal anatomic features in real time, possibly improving safety and precision during surgery. By employing convolutional neural networks (CNNs) for retina vessel segmentation, a retinal coordinate system can be created that allows pre-operative images of capillary non-perfusion or retinal breaks to be digitally aligned and overlayed upon the surgical field in real time. Such technology may be useful in assuring thorough laser treatment of capillary non-perfusion or in using pre-operative optical coherence tomography (OCT) to guide macular surgery when microscope-integrated OCT (MIOCT) is not available.

View Article and Find Full Text PDF

Rapid Acquisition of High-Pixel Fluorescence Lifetime Images of Living Cells via Image Reconstruction Based on Edge-Preserving Interpolation.

Biosensors (Basel)

January 2025

State Key Laboratory of Radio Frequency Heterogeneous Integration (Shenzhen University), Key Laboratory of Optoelectronic Devices and Systems of Ministry of Education and Guangdong Province, College of Physics and Optoelectronic Engineering, Shenzhen University, Shenzhen 518060, China.

Fluorescence lifetime imaging (FLIM) has established itself as a pivotal tool for investigating biological processes within living cells. However, the extensive imaging duration necessary to accumulate sufficient photons for accurate fluorescence lifetime calculations poses a significant obstacle to achieving high-resolution monitoring of cellular dynamics. In this study, we introduce an image reconstruction method based on the edge-preserving interpolation method (EPIM), which transforms rapidly acquired low-resolution FLIM data into high-pixel images, thereby eliminating the need for extended acquisition times.

View Article and Find Full Text PDF

This study presents the characterization of a novel multilayered three-dimensional (3D) polymer exhibiting aggregation-induced emission (AIE) properties when excited at a low wavelength of 280 nm. Utilizing fluorescence spectroscopy, we demonstrate that the polymer displays a marked enhancement in luminescence upon aggregation, a characteristic behavior that distinguishes AIE-active materials from conventional fluorophores. Furthermore, we explore the potential application of this multilayered 3D polymer as a fluorescent probe for the selective detection of specified metal ions.

View Article and Find Full Text PDF

Self-absorption effect in soft X-ray emission spectra utilized for bandgap evaluation of semiconductors.

Microscopy (Oxf)

January 2025

Institute of Multidisciplinary Research for Advanced Materials, Tohoku University, 2-1-1 Katahira, Aoba-ku, Sendai, Miyagi 980-8577, Japan.

The self-absorption effects observed in the background intensity just above the Si L-emission spectra of Si and β-Si3N4, and the C K-emission spectra of diamond and graphite were examined. Based on comparisons with reported results, the energy positions of absorption edges-representing the bottom of conduction bands (CB)-were assigned. The self-absorption profiles in the background intensities were consistent with previously reported data.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!