Background: BioGlue Surgical Adhesive (CryoLife, Inc, Kennesaw, GA) is used to reinforce anastomoses during cardiovascular operations. Previous reports have raised concerns that adhesives may leak through suture-line needle holes and that resulting intraluminal glue may embolize. The purpose of this study was to determine if BioGlue leaks through anastomotic needle holes in aortic tissue and two common prosthetic graft materials.
Methods: Polypropylene suture was used to create end-to-end anastomoses in gelatin-sealed woven polyester grafts (n = 45), expanded polytetrafluoroethylene (ePTFE) grafts (n = 45), and fresh porcine aortas (n = 45). An additional 45 anastomoses were created in ePTFE grafts using ePTFE sutures. The outer surface of each anastomosis was covered with BioGlue. Anastomoses underwent inspection with direct magnification or histology.
Results: BioGlue leaked through needle holes and into the lumen in 10% of anastomoses (18 of 180). Leaks were significantly more common in fresh aorta (10 of 45, 22%) than in prosthetic grafts (8 of 135, 6%; p = 0.003). Suture size did not significantly affect the incidence of leak. The use of ePTFE sutures did not eliminate BioGlue leakage. Prosthetic graft leaks created discreet round adhesive particles. In contrast, aortic tissue leaks resulted in thin, friable flakes of glue extending along the intimal surface. Aortic histology confirmed that BioGlue reached the vessel lumen via the suture channels.
Conclusions: BioGlue leaked through the needle holes in fresh aortic tissue and prosthetic grafts. Intraluminal adhesive particles were easily dislodged, supporting concerns regarding embolization. The potential for adhesive embolization should be a factor when considering the relative risks and benefits of using BioGlue.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.athoracsur.2005.02.004 | DOI Listing |
Aesthetic Plast Surg
January 2025
Department of Body Contouring and Fat grafting Center, Plastic Surgery Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, No 33 Badachu Road, Shijingshan District, 100144, Beijing, People's Republic of China.
Background: Superficial autologous fat grafting is widely used for facial rejuvenation and regenerative treatments, yet its retention rate remains unpredictable. Enhancing the concentration of adipose-derived stem cells (ASCs) has been shown to improve graft retention. Mechanical fat processing techniques can reduce fat granules size for superficial injection as well as elevate the concentration of ASCs.
View Article and Find Full Text PDFDrug Deliv Transl Res
December 2024
Department of Bioscience and Bioinformatics, Kyushu Institute of Technology, 680-4 Kawazu, Iizuka, Fukuoka, 820-8502, Japan.
Int J Pharm
January 2025
Department of Biomaterials Science (BK21 Four Program), Life and Industry Convergence Institute, Pusan National University, Miryang 50463, Republic of Korea. Electronic address:
With increasing clinical demands for painless and easy administration of medications, such as for hair loss, microneedles (MNs) have been widely exploited for facilitating drug permeation in a minimally invasive manner. However, precise dose control and long-term drug delivery without the infection risk through punctured holes have remained unresolved. Herein, we developed swellable microneedles (MNs) with an air-pocket structure, enabling shear-induced implantation inside the skin.
View Article and Find Full Text PDFJ Vitreoretin Dis
October 2024
Kyorin Eye Center, Kyorin University School of Medicine, Tokyo, Japan.
To present a surgical technique that combines an inverted internal limiting membrane (ILM) flap and subretinal injection of balanced salt solution for the closure of chronic macular holes (MH). Two cases are presented. A 74-year-old man and a 63-year-old woman presented with decreased visual acuity (VA) of 20/600 and 20/200, respectively, resulting from a chronic MH.
View Article and Find Full Text PDFBiosensors (Basel)
September 2024
Department of Molecular Compounds Physics, State Research Institute Center for Physical Sciences and Technology, LT-10257 Vilnius, Lithuania.
Black silicon was discovered by accident and considered an undesirable by-product of the silicon industry. A highly modified surface, consisting of pyramids, needles, holes, pillars, etc., provides high light absorption from the UV to the NIR range and gives black silicon its color-matte black.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!