Coronary calcium measurements: effect of CT scanner type and calcium measure on rescan reproducibility--MESA study.

Radiology

Div of Cardiology, Los Angeles Biomedical Research Inst, 1124 W Carson St, Bldg E-5, Torrance, CA 90502, USA.

Published: August 2005

Purpose: To evaluate the effect of scanner type and calcium measure on the reproducibility of calcium measurements.

Materials And Methods: This investigation was approved by the institutional review boards of each study site and by the Institutional Review Board of the Los Angeles Biomedical Research Institute. Informed consent for scanning and participation was obtained from all participants. The study was Health Insurance Portability and Accountability Act compliant. The Multi-Ethnic Study of Atherosclerosis (MESA) is a multicenter observational study of 6814 participants undergoing demographic, risk factor, and subclinical disease evaluations. Coronary artery calcium was measured by using duplicate CT scans. Three study centers used electron-beam computed tomography (CT), and three used multi-detector row CT. Coronary artery calcium was detected in 3355 participants. Three calcium measurement methods-Agatston score, calcium volume, and interpolated volume score-were evaluated. Mean absolute differences between calcium measures on scans 1 and 2, excluding cases for which both scans had a measure of zero, was modeled by using linear regression to compare reproducibility between scanner types. A repeated measures analysis of variance test was used to compare reproducibility across calcium measures, with mean percentage absolute difference as the outcome measure. Rescan reproducibility in relation to misregistrations, noise, and motion artifacts was also examined. Variables were log transformed to create a more normal distribution.

Results: Concordance for presence of calcium between duplicate scans was high and similar for both electron-beam and multi-detector row CT (96%, kappa = 0.92). Mean absolute difference between calcium scores for the two scans was 15.8 for electron-beam and 16.9 for multi-detector row CT scanners (P = .06). Mean relative differences were 20.1 for Agatston score, 18.3 for calcium volume, and 18.3 for interpolated volume score (P < .01). Reproducibility was lower for scans with versus those without image misregistrations or motion artifacts (P < .01 for both).

Conclusion: Electron-beam and multi-detector row CT scanners have equivalent reproducibility for measuring coronary artery calcium. Calcium volumes and interpolated volume scores are slightly more reproducible than Agatston scores. Reproducibility is lower for scans with misregistrations or motion artifacts.

Download full-text PDF

Source
http://dx.doi.org/10.1148/radiol.2362040513DOI Listing

Publication Analysis

Top Keywords

multi-detector row
16
calcium
14
coronary artery
12
artery calcium
12
interpolated volume
12
motion artifacts
12
scanner type
8
type calcium
8
calcium measure
8
measure rescan
8

Similar Publications

Improvement of image quality of dentomaxillofacial region in ultra-high resolution computed tomography: A phantom study.

Dentomaxillofac Radiol

November 2024

Department of Clinical Radiology, Graduate School of Medical Sciences, Kyushu University, 3‑1‑1 Maidashi, Higashi‑ku, Fukuoka, 812‑8582, Japan.

Objectives: The purpose of this study was to compare the image quality of ultra-high-resolution computed tomography (U-HRCT) with that of conventional multi-detector row CT (convCT) and demonstrate its usefulness in the dentomaxillofacial region.

Methods: Phantoms were helically scanned with U-HRCT and convCT scanners using clinical protocols. In U-HRCT, phantoms were scanned in super-high-resolution (SHR) mode, and hybrid iterative reconstruction (HIR) and filtered-back projection (FBP) techniques were performed using a bone kernel (FC81).

View Article and Find Full Text PDF

Background: Peripheral arterial disease (PAD) of the lower limbs is caused by atherosclerotic occlusive disease in which narrowing of arteries reduces blood flow to the lower limbs. PAD is common; it is estimated to affect 236 million individuals worldwide. Advanced age, smoking, hypertension, diabetes and concomitant cardiovascular disease are common factors associated with increased risk of PAD.

View Article and Find Full Text PDF

A retrieval of relevant literature on hepatic nodular lesions, gastric cancer (GC), and Crohn's disease (CD) was conducted from Chinese and English databases. Meta-analysis was performed using Review Manager 5.4 software and the MIDAS package in Stata 18.

View Article and Find Full Text PDF

Objectives: To assess the relationships between longitudinal fluoride intakes and bone densitometry outcomes in young adults.

Methods: Data were analyzed from the Iowa Fluoride Study and Iowa Bone Development Study, which followed 1,882 infants from birth in 1992. Daily fluoride intakes were assessed using detailed questionnaires sent every 1.

View Article and Find Full Text PDF

Depiction rate of feeding arteries of renal cell carcinoma on four-dimensional computed tomography angiography.

Jpn J Radiol

June 2024

Department of Radiology, Graduate School of Medicine, Dentistry and Pharmaceutical Sciences, Okayama University, 2-5-1 Shikata-Cho, Kita-Ku, Okayama, 700-8558, Japan.

Purpose: To retrospectively evaluate the depiction rate of feeding arteries in biopsy-proven clear cell renal cell carcinoma (CCRCC) on four-dimensional computed tomography angiography (4D-CTA) images.

Materials And Methods: This study included 22 patients with 22 CCRCC and 30 feeding arteries treated with transcatheter renal artery embolization. The depiction rate of the feeding arteries on preprocedural 4D-CTA was evaluated.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!