The performance of poly-methyl pentene (PMP) oxygenators (Medos Hilite 7000LT) was compared with that of silicone membrane (SM) oxygenators (Medtronic 1-4500-2A) for adult extracorporeal membrane oxygenation (ECMO). Forty consecutive patients were selected retrospectively pre- and post-introduction of PMP oxygenators. They were selected according to the dates they received ECMO and were separated into two equal groups with similar backgrounds. The flow path resistance, gas and heat exchange efficiency, consumption of coagulation factors and platelets, blood transfusion requirements, and incidence of clots for each oxygenator type was assessed. Adult PMP oxygenators showed lower blood path resistance than SM oxygenators. However, lower consumption of blood products in these oxygenators was a direct result of their smaller surface area and heparin coated design, reducing contact activation of coagulation factors. These oxygenators are noticeably smaller, require lower priming volumes, and have better gas exchange capability than SM oxygenators. They showed greater stability and preservation of coagulation factors and platelets compared with SM oxygenators. They also had the advantage of a functioning integrated heat exchanger. Using a single PMP oxygenator in the first instance may be adequate for the majority of patients and would significantly reduce red blood cell consumption during ECMO.

Download full-text PDF

Source
http://dx.doi.org/10.1097/01.mat.0000159741.33681.f1DOI Listing

Publication Analysis

Top Keywords

pmp oxygenators
12
coagulation factors
12
oxygenators
10
poly-methyl pentene
8
gas exchange
8
exchange capability
8
transfusion requirements
8
adult extracorporeal
8
extracorporeal membrane
8
membrane oxygenation
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!