Development of activity standard for 90 Y microspheres.

Appl Radiat Isot

Australian Nuclear Science and Technology Organisation, New Illawarra Road, Lucas Heights, NSW 2234, Australia.

Published: August 2005

(90)Y microspheres are important therapeutic radiopharmaceuticals used in the treatment of liver cancer through a process known as selective internal radiation therapy. SIR-spheres is a radiopharmaceutical product that is comprised of (90)Y microspheres suspended in sterile, pyrogen-free water for injection into patients. It is necessary to establish for the SIR-spheres production the capability of accurately measuring the activity of this product to a traceable national measurement standard. An activity standard for SIR-spheres was developed from a standard for (90)Y solution, employing a highly quantifiable chemical digestion process. Calibration factors for the manufacturer's ionisation chambers were determined for 1 and 5 ml of the SIR-spheres product placed in Wheaton vials, for both 34% and 44% of (90)Y microsphere concentration.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.apradiso.2005.02.002DOI Listing