Comparison of erythrocyte sedimentation rate measurement by the automated SEDIsystem and conventional Westergren method using the Bland and Altman statistical method.

Med Princ Pract

Department of Medical Laboratory Sciences, Faculty of Allied Health Sciences, Kuwait University, Sulaibekhat, Kuwait.

Published: February 2006

Objectives: To compare the performance of SEDIsystem(TM), a fully automated analyzer for the measurement of the erythrocyte sedimentation rate (ESR), with the manual Westergren method.

Materials And Methods: Both methods were applied to 150 randomly selected subjects. The linear regression and Bland and Altman data analysis methods were used to measure the agreement between the automated and manual methods.

Results: The regression analysis showed a good correlation between the two methods (r=0.91). The Bland and Altman data analysis showed no systematic bias (95% confidence interval for mean difference); however, limits of agreement were between 11.52 and -37.88. This indicates that ESR values measured by the SEDIsystem may be 11.52 mm/h above or 37.88 mm/h below the reference method. A greater scatter of data was also observed with abnormally high (>25 mm/h) ESR results (mean of difference=-21.4 and limits of agreement=-45.2 and 2.26) compared with normal (<25 mm/h) readings (mean of difference=-3.9 and limits of agreement=-13.5 and 5.7).

Conclusion: The Bland and Altman statistical analysis showed a wide degree of scatter between results obtained by the two ESR techniques that was not clearly demonstrated using the linear regression analysis. The automated system was found to underestimate ESR with the Bland and Altman statistical analysis, and therefore a correction factor is recommended.

Download full-text PDF

Source
http://dx.doi.org/10.1159/000085742DOI Listing

Publication Analysis

Top Keywords

bland altman
12
erythrocyte sedimentation
8
sedimentation rate
8
altman data
8
data analysis
8
comparison erythrocyte
4
rate measurement
4
measurement automated
4
automated sedisystem
4
sedisystem conventional
4

Similar Publications

Purpose: To assess the effects of bladder fullness and lower uterine contractions ultrasound on transabdominal and transvaginal cervical length measurements at the mid-trimester fetal anomaly scan (FAS).

Methods: Transabdominal and transvaginal cervical length measurements from 925 mid-trimester FAS examinations were retrospectively analysed. Images were assessed for lower uterine contraction and bladder fullness using a novel qualitative assessment.

View Article and Find Full Text PDF

The 100-m sprint is one of the track events, and the pace of the runner can be measured using a variety of tools, such as a hand stopwatch, timing gate, laser device, radar device, photocell timing, etc. The data measured is the mean travel time. Nonetheless, monitoring an individual sprinter's instantaneous speed tracking is essential for assisting staff trainers in developing an appropriate training schedule for the individual sprinter.

View Article and Find Full Text PDF

Measurements of Objective Cyclotorsion in a Population of Healthy Children.

J Ophthalmol

December 2024

Department of Ophthalmology, Chang Gung Memorial Hospital, Linkou Medical Center, Taoyuan, Taiwan.

This study aimed to compare ocular torsion measurements to investigate normative objective cyclotorsion values in a population of healthy, full-term and preterm children. The participants enrolled in this study had an age range of 3-12 years and were divided into two groups, full-term (gestational age (GA) > 37 weeks) and preterm without retinopathy of prematurity (ROP) (GA ≤ 37 weeks). The disc-center-fovea angle (DFA) was used to evaluate ocular torsion using two different imaging modalities: optical coherence tomography (OCT) with a 55-degree field of view (FV) and conventional fundus photography (CFP) with a 45-degree FV.

View Article and Find Full Text PDF

PET/CT for perforator identification in deep inferior epigastric perforator flap harvest.

Am J Otolaryngol

December 2024

Department of Otolaryngology-Head and Neck Surgery, Indiana University School of Medicine, Indianapolis, IN, USA; Indiana University School of Medicine, Indianapolis, IN, USA. Electronic address:

Background: CT angiography (CTA) is used for preoperative localization in deep inferior epigastric perforator (DIEP) flaps, but is an additional costly study that involves contrast and radiation exposure. Many patients with head and neck cancer already undergo PET/CT. We investigated if PET/CT could be used to preoperatively localize perforators and if this corresponded with the intraoperative location.

View Article and Find Full Text PDF

Verification imaging in prostate MR-only radiotherapy: Are fiducial markers necessary?

Radiography (Lond)

December 2024

Newcastle Upon Tyne Hospitals NHS Foundation Trust, Northern Centre for Cancer Care, Newcastle Upon Tyne, United Kingdom; Newcastle University, Translational and Clinical Research Institute, Newcastle Upon Tyne, United Kingdom.

Purpose/objective: MR-only radiotherapy planning exploits the benefits of MRI soft-tissue delineation, whilst negating the registration inaccuracies caused by MRI CT fusion. Fiducial markers have conventionally been used in prostate radiotherapy to reduce on-treatment image matching variability. However, this is an invasive procedure for the patient, and presents technical difficulties in an MR-only pathway as fiducial markers are difficult to visualise on MRI.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!