Virtual micro MLC commissioning.

J Appl Clin Med Phys

Medical Physics Department, Allan Blair Cancer Centre, Regina, Saskatchewan, S4S 1V4, Canada.

Published: March 2006

The resolution of multileaf collimators (MLCs) is limited by their finite leaf width. A commercial package (HD-270) uses 3D couch translation and leaf adjustments to emulate smaller leaf widths. In this paper, we report on the commissioning of this feature using software testing, dial gauge measurements, and film dosimetry. We also identify a variety of limitations: software bugs and truncation artifacts, MLC leaf positioning uncertainties (random variations, systematic gantry dependence and backlash), and uncertainties in couch positioning. These reduce the capabilities of this implementation below that achievable theoretically.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5723476PMC
http://dx.doi.org/10.1120/jacmp.v6i2.2032DOI Listing

Publication Analysis

Top Keywords

virtual micro
4
micro mlc
4
mlc commissioning
4
commissioning resolution
4
resolution multileaf
4
multileaf collimators
4
collimators mlcs
4
mlcs limited
4
limited finite
4
leaf
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!