Effective comparison of two auto-CPAP devices for treatment of obstructive sleep apnea based on polysomnographic evaluation.

Auris Nasus Larynx

Department of Otorhinolaryngology, Graduate School of Medical Sciences, Kyushu University, Fukuoka 812-8582, Japan; Sendai Snoring Research Centre, Miyoshi ENT Clinic, Sendai 981-3133, Japan.

Published: September 2005

Background: Automatic continuous positive airway pressure (auto-CPAP) machines differ mainly in algorithms used for respiratory event detection and pressure control. The auto-CPAP machines operated by novel algorithms are expected to have better performance than the earlier ones in the treatment of obstructive sleep apnea syndrome (OSAS).

Objectives: The purpose of this study was to determine the therapeutic characteristics between two different auto-CPAP devices, i.e., the third-generation flow-based (f-APAP) and the second-generation vibration-based (v-APAP) machines, during the first night treatment of OSAS.

Methods: We retrospectively reviewed the polysomnography (PSG) recordings of 43 OSAS patients who were initially performed an overnight diagnostic PSG to confirm the disease and afterwards received the first night auto-CPAP treatment with using either the f-APAP (n=22) or v-APAP (n=21) device under another PSG evaluation.

Results: There were 13.6% and 61.9% patients who remained a residual apnea/hypopnea index more than 5 during the f-APAP and v-APAP application, respectively (P<0.005). The f-APAP was more effective than the v-APAP in reducing apnea/hypopnea index (P=0.003), hypopnea index (P=0.023) and apnea index (P=0.007), improving the lowest oxygen saturation index (P=0.007) and shortening stage 1 sleep (P=0.016). However, the f-APAP was less sufficient than the v-APAP in reducing arousal/awakening index (P=0.02).

Conclusion: These findings suggest that the f-APAP works better than the v-APAP in abolishing breathing abnormities in the treatment of OSAS; however, the f-APAP device might still have some potential limitations in the clinical application.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.anl.2005.03.007DOI Listing

Publication Analysis

Top Keywords

auto-cpap devices
8
treatment obstructive
8
obstructive sleep
8
sleep apnea
8
auto-cpap machines
8
auto-cpap
5
effective comparison
4
comparison auto-cpap
4
treatment
4
devices treatment
4

Similar Publications

Background: Obstructive sleep apnea (OSA) is a common disorder characterized by recurrent upper airway collapse during sleep, leading to neurological and cardiovascular adverse effects.

Objective: The present study aimed to explore seasonal compliance with Continuous Positive Airway Pressure (CPAP) therapy among OSA patients in Greece.

Methods: Data from 954 OSA patients using CPAP devices were collected during summer and winter months.

View Article and Find Full Text PDF

Sleep electroencephalogram (EEG) signals likely encode brain health information that may identify individuals at high risk for age-related brain diseases. Here, we evaluate the correlation of a previously proposed brain age biomarker, the "brain age index" (BAI), with cognitive test scores and use machine learning to develop and validate a series of new sleep EEG-based indices, termed "sleep cognitive indices" (SCIs), that are directly optimized to correlate with specific cognitive scores. Three overarching cognitive processes were examined: total, fluid (a measure of cognitive processes involved in reasoning-based problem solving and susceptible to aging and neuropathology), and crystallized cognition (a measure of cognitive processes involved in applying acquired knowledge toward problem-solving).

View Article and Find Full Text PDF

Update on the treatment of idiopathic hypersomnia: Progress, challenges, and expert opinion.

Sleep Med Rev

June 2023

National Reference Network for Orphan Diseases: Narcolepsy and Rare Hypersomnias, Paris, France; Sleep and Wake Disorders Centre, Department of Neurology, Gui de Chauliac Hospital, Montpellier, France; University of Montpellier, INSERM Institute Neuroscience Montpellier (INM), Montpellier, France.

Idiopathic hypersomnia is a central hypersomnolence disorder of unknown origin characterized by excessive daytime sleepiness despite normal or long sleep time, and frequent severe sleep inertia. Management strategies have been largely derived from expert consensus, due to a lack of disease-specific assessments and reliance on case series and rare randomized controlled studies. Guidelines recommend treatment with off-label medications.

View Article and Find Full Text PDF

Obstructive sleep apnea syndrome (OSAS) is a common sleep-related breathing disorder where precise treatment assessment is of high importance. We aimed to validate an automatic algorithm of the auto-CPAP device and reveal polygraph usefulness in the OSAS diagnosis and treatment of outpatients. One hundred patients with moderate OSAS, severe OSAS, and excessive daytime sleepiness qualified for CPAP treatment were included.

View Article and Find Full Text PDF

Purpose: Sleep-disordered breathing may be induced by, exacerbate, or complicate recovery from critical illness. Disordered breathing during sleep, which itself is often fragmented, can go unrecognized in the intensive care unit (ICU). The objective of this study was to investigate the prevalence, severity, and risk factors of sleep-disordered breathing in ICU patients using a single respiratory belt and oxygen saturation signals.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!