Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Study Objectives: Transcutaneous CO(2) pressure (Ptcco(2)) and transcutaneous O(2) pressure (Ptco(2)) measurements are routinely used in pediatric ICUs in order to avoid serial arterial punctures. The aim of this study was to determine the value of Ptcco(2) assessment during the evaluation of home ventilation in 12 adult patients with COPD or restrictive respiratory failure in the stable state (mean [+/- SD] basal Paco(2), 48.8 +/- 8.3 mm Hg) who were treated by mask or tracheotomy-mediated ventilation.
Methods: After radial catheter insertion, patients were instructed to breathe spontaneously for 40 min and then to receive ventilation for 40 min according to their individual home ventilation modalities. An in vivo calibration was performed in the initial stage of the study in order to optimize the arterial Pco(2) and Ptcco(2) values. Every 5 min, transcutaneous measurements were performed and simultaneously compared with arterial values.
Measurements And Results: Ptcco(2) and Ptco(2) were correlated with arterial values (p < 0.0001) except for Paco(2) values of > 56 mm Hg and Pao(2) values of > 115 mm Hg. During ventilation, Paco(2) decreased >or= 4 mm Hg in seven patients. Ptcco(2) variations recorded during consecutive 5-min periods while the patient received mechanical ventilation were well correlated with the arterial variations (p = 0.0033), with a delay of < 5 min.
Conclusion: Ptcco(2) values and variations accurately reflected Paco(2) values and variations during mechanical ventilation. However, the accuracy of these data seems to be restricted to patients with Paco(2) values of < 56 mm Hg.