Microtrap modulated flame ionization detector for on-line monitoring of methane.

J Chromatogr A

Analytical and Environmental/Trace Analysis Research Unit, Department of Chemistry, Faculty of Science, Prince of Songkla University, Hat Yai, Songkhla 90112, Thailand.

Published: April 2005

A microtrap for on-line analysis of trace methane was developed. Silicosteel tubing, 15 cm long, 1.02 mm i.d. and 1.59 mm o.d. was packed with Carbosphere 80/100 mesh and placed between the standard gas source and the flame ionization detector (FID) to be used as a preconcentrator and an injector. The sample stream was passed continuously through the microtrap that was heated at a fixed interval by an electric pulse. To achieve the highest adsorption and desorption efficiency, parameters affecting the response of the system were optimized, and adsorption temperature in the range of 25 to -50 degrees C was studied. The response from the microtrap as compared to that from direct-flame ionization detector was enhanced by as much as 260 times. The on-line microtrap system showed good linearity (r2 > 0.99), low detection limit (28.3 ppbv) and good long term stability (relative standard deviation, RSD of less than 5.0%). Therefore, this simple device is suitable for on-line analysis of trace methane and similar small molecules in the environment as in high purity gases.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.chroma.2005.03.041DOI Listing

Publication Analysis

Top Keywords

ionization detector
12
flame ionization
8
on-line analysis
8
analysis trace
8
trace methane
8
microtrap
5
microtrap modulated
4
modulated flame
4
on-line
4
detector on-line
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!