On concomitant gradients in low-field MRI.

J Magn Reson

Los Alamos National Laboratory, Biological and Quantum Physics Group, Los Alamos, NM 87545, USA.

Published: July 2005

AI Article Synopsis

Article Abstract

Growing interest in magnetic resonance imaging (MRI) at ultra-low magnetic fields (ULF, approximately muT fields) has been motivated by several advantages over its counterparts at higher magnetic fields. These include narrow line widths, the possibility of novel imaging schemes, reduced imaging artifacts from susceptibility variations within a sample, and reduced system cost and complexity. In addition, ULF NMR/MRI with superconducting quantum interference devices is compatible with simultaneous measurements of biomagnetic signals, a capability conventional systems cannot offer. Acquisition of MRI at ULF must, however, account for concomitant gradients that would otherwise result in severe image distortions. In this paper, we introduce the general theoretical framework that describes concomitant gradients, explain why such gradients are more problematic at low field, and present possible approaches to correct for these unavoidable gradients in the context of a non-slice-selective MRI protocol.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jmr.2005.03.015DOI Listing

Publication Analysis

Top Keywords

concomitant gradients
12
magnetic fields
8
gradients low-field
4
mri
4
low-field mri
4
mri growing
4
growing interest
4
interest magnetic
4
magnetic resonance
4
resonance imaging
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!