Background And Purpose: Vascular control is crucial during laparoscopic ligation of the renal vessels. Intracorporeal suture ligation has been replaced by easier methods, such as specialized clip appliers and automatic stapling devices; nevertheless, the optimal application and margin of safety of such devices have yet to be determined. We sought to address this question by measuring the bursting strength of arteries ligated with several standard devices.
Materials And Methods: One end of an adult porcine artery (3-7-mm diameter) was occluded with a titanium clip, self-locking polymer clip, or laparoscopic linear cutting stapler. Comparisons were made with one or two clips and with different distal cuff lengths (i.e., flush or 2 mm). The open end was secured to a pulsatile infusion pump. Leak/failure pressures were measured using a digital barometer.
Results: The mean bursting pressures for the clips were above physiologic arterial pressures (1220-1500 mm Hg). However, the vessels closed with the stapler leaked at a lower mean pressure (262 mm Hg). Failure of titanium or self-locking polymer clips was the result of vessel retraction into and behind the clip, while staple- line leakage occurred between individual staples. Bursting pressures with the titanium and self-locking polymer clips was unaffected by the number of clips or length of vascular cuff.
Conclusions: All tested methods of vascular control performed well at physiologic pressures, suggesting that safety is not increased with traditional maneuvers such as additional clips or longer cuff length.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1089/end.2005.19.307 | DOI Listing |
Adv Sci (Weinh)
June 2024
Engineering and Applied Science, California Institute of Technology, 1200 E. California Blvd., Pasadena, CA, 91125, USA.
Multi-stable elements are commonly employed to design reconfigurable and adaptive structures, because they enable large and reversible shape changes in response to changing loads, while simultaneously allowing self-locking capabilities. However, existing multi-stable structures have properties that depend on their initial design and cannot be tailored post-fabrication. Here, a novel design approach is presented that combines multi-stable structures with two-way shape memory polymers.
View Article and Find Full Text PDFVet Surg
February 2024
JT Vaughan Large Animal Teaching Hospital, Auburn University, Auburn, Alabama, USA.
Objective: The aim of the study was to evaluate the strength and size of the double forwarder (DF) knot in 2 and 3 USP polyglactin 910 when used to form a ligature and to compare the knot holding capacity (KHC), size and weight of the DF knot to surgeon's (SU) and square (SQ) knots with varying numbers of throws.
Study Design: Laboratory study.
Study Population: Knotted suture.
Acta Vet Scand
November 2023
Department of Clinical Sciences, Faculty of Veterinary Medicine and Animal Science, Swedish University of Agricultural Sciences (SLU), Box 7054, Uppsala, S-750 07, Sweden.
Background: A 3D printed self-locking device made of polydioxanone (PDO) was developed to facilitate a standardized ligation technique. The subcutaneous tissue reaction to the device was evaluated after implantation in ten horses of mixed age, sex and breed and compared to loops of poly(lactic-co-glycolic acid) (PLGA). In two of the horses, the implants were removed before closing the skin.
View Article and Find Full Text PDFBiomater Sci
September 2023
Key Laboratory of Textile Science & Technology, Ministry of Education, College of Textiles, Donghua University, Shanghai, China.
Polypropylene (PP) sling implantation is the most commonly performed procedure for women with stress urinary incontinence (SUI). However, concerns have arisen regarding complications caused by slings, including the common issue of erosion, which can be attributed to various factors such as the body's response and bacterial contamination. To address these concerns, we have developed a rectangular mesh self-locking edge sling with a large pore size and lightweight design.
View Article and Find Full Text PDFMedicina (Kaunas)
March 2023
Faculty of Medicine, Vilnius University, LT-03101 Vilnius, Lithuania.
Closure of the appendix stump is critical to avoid serious postoperative complications. There are a number of options, but the best one has not been identified yet. The purpose of this study is to evaluate the outcomes of appendiceal stump closure using self-locking polymeric clips and endoloops.