Objectives: A protocol using transthoracic echocardiography was designed to diagnose the common malfunctions of patients on chronic support with a left ventricular assist device (LVAD).
Background: Mechanical circulatory support, primarily with a LVAD, is increasingly used for treatment of advanced heart failure as a bridge to transplant and for long-term treatment of heart failure. The LVAD dysfunction is a recognized complication. To date, no studies have defined the role of transthoracic echocardiography in evaluating long-term mechanical complications of chronic LVAD support.
Methods: Transthoracic echocardiography was used in a protocol designed to detect the common types of mechanical malfunction. Patients were followed up with serial echocardiograms, and clinical validations were made with findings from a catheter-based protocol and inspection at the time of cardiac transplant or corrective surgery.
Results: Thirty-two patients with 44 LVADs were followed up during a four-year period using this protocol that correctly identified 11 patients with inflow valve regurgitation, 2 with intermittent inflow conduit obstruction, 1 with severe kinking of the outflow graft, and 9 with new insufficiency of the native aortic valve.
Conclusions: As LVAD use for end-stage heart failure becomes widespread, and durations of support are extended, dysfunction will be increasingly prevalent. Transthoracic echocardiography provides a practical method to accurately identify the causes of mechanical dysfunction with patients on chronic LVAD support.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jacc.2005.01.037 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!