Accurate reporting of personal dose is required by regulation for hospital personnel that work with radioactive material. Pocket dosimeters are commonly used for monitoring this personal dose. We show that operating a cell phone in the vicinity of a pocket dosimeter can introduce large and erroneous readings of the dosimeter. This note reports a systematic study of this electromagnetic interference. We found that simple practical measures are enough to mitigate this problem, such as increasing the distance between the cell phone and the dosimeter or shielding the dosimeter, while maintaining its sensitivity to ionizing radiation, by placing it inside a common anti-static bag.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1088/0031-9155/50/9/N03 | DOI Listing |
Indian J Nucl Med
December 2023
Department of Nuclear Medicine, Postgraduate Institute of Medical Education and Research, Chandigarh, India.
Purpose: The present study aimed to evaluate radiation exposure to staff performing coronary flow reserve (CFR) measurement using N-ammonia.
Materials And Methods: The radiation exposure rate during the administration of N-ammonia for the rest and stress part of the study was noted using an ionization chamber-based calibrated survey monitor. The radiation exposure to persons involved in dispensing radioactivity (D1), administering radioactivity (D2) and monitoring the patient during pharmacological stress (D3) were measured using an energy compensated Si-diode personal pocket dosimeter.
Oral Radiol
April 2024
Department of Oral and Maxillofacial Radiology, Graduate School of Biomedical and Health Sciences, Hiroshima University, Hiroshima, 734-8553, Japan.
Objectives: Low-dose-rate brachytherapy (LDR-BT) with Au grains and Ir pins is an essential treatment option for oral cancer due to its high rate of local control and low invasiveness. However, the radiation exposure of medical radiation workers is concerning. Thus, we aimed to determine the radiation dose delivered to medical radiation workers during LDR-BT using Au grains and Ir pins for oral cancer.
View Article and Find Full Text PDFHealth SA
July 2023
Department of Pharmaceutical Sciences, School of Pharmacy, Sefako Makgatho Health Sciences University, Pretoria, South Africa.
Background: Positron emission tomography/computed tomography (PET/CT) is an imaging modality that combines images from high-energy gamma rays emitted by a positron emitting radiopharmaceutical and those from the CT component. The images are then used in the diagnosis of severe diseases. Procedures with PET radiopharmaceuticals introduce a risk of high occupational radiation exposure to staff handling them.
View Article and Find Full Text PDFCardiovasc Intervent Radiol
June 2023
Department of Radiology, Hyogo Medical University, 1-1 Mukogawa-cho, Nishinomiya, Hyogo Prefecture, 663-8501, Japan.
Purpose: To evaluate, experimentally and clinically, the radioprotective effects of a semicircular X-ray shielding device for operators during CT fluoroscopy-guided IR procedures.
Materials And Methods: During experimentation, the reduction rates of scattered radiation rates from CT fluoroscopy were evaluated using a humanoid phantom. Two shielding device positions were tested: "shielding close to the CT gantry" and "shielding close to the operator".