Aim: To reassess the cut-off value for lamellar body counts (LBs) for fetal lung maturity (FLM) over a 10-year study period.

Patients And Methods: 178 pregnancies were selected under strict inclusion criteria and delivered within 48 h from amniocentesis. FLM was determined by amniotic fluid LBs in centrifuged samples (300 x g for 10 min) in a commercially available Coulter Counter. Cases beyond 37 weeks were excluded.

Results: Mean gestational age was 33.5+/-3.0 weeks at amniocentesis and 33.7+/-3.0 weeks at birth. After reassessing the best compromise between sensitivity and specificity for all cases using the receiver operating characteristic (ROC) procedure, an FLM cut-off value of < or = 22,000/microL was obtained. Diagnostic accuracy (and confidence interval, CI) was: sensitivity, 73% (60.0-83.6%); specificity, 81.7% (CI 73.6-88.1%); positive predictive value, 66.2%; and negative predictive value, 86.0%.

Conclusion: No significant change in FLM cut-off for LBs was found when comparing the value from this study and the results of our earlier report presented in 1996 (< or = 22,000 vs. < or = 20,000/microL), although the new value may be more accurate, since it is based on neonatal outcome with the exclusion of cases in which the diagnosis of FLM is seldom warranted, i.e., > 37 weeks' gestational age.

Download full-text PDF

Source
http://dx.doi.org/10.1515/JPM.2005.029DOI Listing

Publication Analysis

Top Keywords

amniotic fluid
8
lamellar body
8
body counts
8
fetal lung
8
lung maturity
8
gestational age
8
flm cut-off
8
flm
5
fluid lamellar
4
counts determination
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!