Aims: The purpose of this study was to evaluate the repeatability of forearm skin blood flow responses to intradermal injections of acetylcholine (ACh) and endothelin-1 (ET-1) using a double injection technique (DIT) and a laser Doppler imager (LDI) scanner in the human skin microcirculation.
Methods: We used a laser Doppler imager (Moor LDI V3.01) to continuously monitor the change in skin blood flow during intradermal administration of physiological saline (0.9% NaCl), acetylcholine (ACh 10(-7), 10(-8), 10(-9) M) and endothelin-1 (ET-1 10(-14), 10(-16), 10(-18) M) in 10 healthy male subjects. Subjects were examined on 3 different days for assessment of interday and interobserver repeatability. Injections of either drug were randomly placed on different sites of the forearm. Laser Doppler images were collected before and after injection at 2.5 min intervals for 30 min. Data were analysed after the completion of each experiment using Moor Software V.3.01. Results are expressed as changes from baseline in arbitrary perfusion units (PU).
Results: ACh caused a significant vasodilation (P < 0.0001 anova, mean +/- SE: 766 +/- 152 PU, ACh 10(-9) M; 1868 +/- 360 PU, ACh 10(-8) M; 4188 +/- 848 PU, ACh 10(-7) M; mean of days 1 and 2, n = 10), and ET-1 induced a significant vasoconstrictive response (P < 0.0001 anova, -421 +/- 83 PU, ET-1 10(-18) M; -553 +/- 66 PU, ET-1 10(-16) M; -936 +/- 90 PU, ET-1 10(-14) M; mean of days 1 and 2, n = 10). There was no difference on the response to either drug on repeated days. Bland-Altman analyses showed a close agreement of responses between days with repeatability coefficients of 1625.4 PU for ACh, and 386.0 PU for ET-1 (95% CI: ACh, -1438 to 1747 PU, ET-1, -399 to 358 PU) and between observers with repeatability coefficients of 1057.2 PU for ACh and 255.8 PU for ET-1 (95% CI: ACh, -1024 to 1048 PU, ET-1, -252 to 249 PU). The variability between these responses was independent of average flux values for both ACh and ET-1. There was a significant correlation between responses measured in the same site, in the same individual on two different days by the same observer (ACh, r = 0.94, P < 0.0001; ET-1, r = 0.90, P < 0.0006), and between responses measured by two different observers (ACh, r = 0.94, P < 0.0001; ET-1, r = 0.91, P < 0.0003).
Conclusion: We have shown that interday and intraobserver responses to intradermal injections of ET-1 and ACh, assessed using the DIT in combination with an LDI scanner, exhibited good reproducibility and may be a useful tool for studying the skin microcirculation in vivo.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC1884840 | PMC |
http://dx.doi.org/10.1111/j.1365-2125.2004.02344.x | DOI Listing |
Langmuir
January 2025
Cain Department of Chemical Engineering, Louisiana State University, Baton Rouge, Louisiana 70803, United States.
The self-assembly of nanoparticle colloids into large-area monolayers with long-range order is a grand challenge in nanotechnology. Using acoustic energy, i.e.
View Article and Find Full Text PDFJ Biophotonics
January 2025
Research and Development Center of Biomedical Photonics, Orel State University, Orel, Russia.
The work is devoted to the study of the physiological variability of the microcirculatory-tissue system (MTS) parameters under normal conditions and during functional tests. The results were obtained in vivo using multimodal wearable analyzers implementing methods of laser Doppler flowmetry and fluorescence spectroscopy. Comprehensive data analysis and calculation of the coefficients of variation of the MTS parameters of the human body for various topographic and anatomical areas of the skin were carried out.
View Article and Find Full Text PDFMid-infrared dual-comb spectroscopy offers significant advantages by combining the high sensitivity of mid-infrared spectroscopy with the high spectral resolution and rapid acquisition of the dual-comb method. However, its effective resolution, constrained by the inherent comb line spacing, hinders its ability to resolve narrow absorption features, common in critical applications such as sub-Doppler spectroscopy, low-pressure gas analysis, and construction of the atmospheric profile. To address this challenge, we present a synchronous offset frequency tuning method for the mid-infrared dual-comb system to improve effective resolution far beyond comb line spacing.
View Article and Find Full Text PDFWe present a dual isotope magneto-optical trap (MOT), simultaneous sub-Doppler laser cooling, and magnetic trapping of a spin-polarized K-K Bose-Fermi mixture realized in a single-chamber setup with an unenriched potassium dispenser as the source of atoms. We are able to magnetically confine more than 2.2 × 10 fermions ( = 9/2 , = 9/2) and 1.
View Article and Find Full Text PDFJ Cardiothorac Surg
January 2025
Department of Anesthesiology, Zhongda hospital, Southeast University, No. 87 Dingjiaqiao, Nanjing City, 210009, Jiangsu Province, China.
Monitoring perioperative tissue perfusion is crucial in clinical anesthesia to protect organs and ensure patient safety. Indicators like hemodynamic parameters, tissue metabolism, and microcirculation markers are used for assessment. Studies show intraoperative hypotension negatively impacts outcomes, though blood pressure alone may not reflect tissue perfusion accurately.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!