Background: The effect of feeding tube size and port configuration on the ability to measure gastric residual volume (GRV) is poorly understood. In addition, there is confusion about the need to measure GRVs during feedings into the small bowel. This study sought to (1) compare the volume of gastric contents obtained from small-diameter feeding tubes and large-diameter sump tubes concurrently positioned in the stomach and (2) describe the distribution of GRVs during small-bowel feedings.

Methods: For the first objective, GRV measurements were made from 10-Fr tubes (n = 645) and 14-Fr or 18-Fr sump tubes (n = 645) concurrently present in 62 critically ill patients. Sixty-milliliter syringes were used to measure GRVs from the 10-Fr tubes; the fluid was returned to the stomach and measurements were repeated from the large-diameter sump tubes. To address the second research objective, 890 GRV measurements were made from 14-Fr or 18-Fr gastric sump tubes (not connected to suction) in 75 critically ill patients who were receiving small-bowel feedings.

Results: When GRVs were >50 mL, a linear regression equation indicated that volumes obtained from the large-diameter sump tubes were about 1.5 times greater than those obtained from the small-diameter tubes concurrently present in the stomach, p < .001. Gastric volumes > or =100 mL were found in 11.6% of the 890 measurements made in patients receiving small-bowel feedings; volumes > or =150 mL were found in 5.4% of the measurements.

Conclusions: The findings suggest that GRVs obtained from large-diameter sump tubes are about 1.5 times greater than those obtained from 10-Fr tubes. Large GRVs occur in at least 5% of patients receiving postpyloric feedings.

Download full-text PDF

Source
http://dx.doi.org/10.1177/0148607105029003192DOI Listing

Publication Analysis

Top Keywords

sump tubes
24
large-diameter sump
16
10-fr tubes
12
patients receiving
12
tubes
11
residual volume
8
measure grvs
8
tubes concurrently
8
grv measurements
8
tubes 645
8

Similar Publications

Background: Second-generation supraglottic airway devices (SGAs) are pivotal in airway management, with the ability to accommodate gastric tube (GT) insertion. However, research on GT insertion with SGAs under controlled conditions is limited. This study aimed to evaluate the GT insertion performance of SGAs using a manikin.

View Article and Find Full Text PDF

16 Fr Salem Sump™ tubes have special features to facilitate suction drainage of the stomach, including a second lumen for air venting. These tubes are also commonly used to deliver enteral nutrition and medications to intensive care unit (ICU) patients, but we found no previous research to justify this practice. Because of the unused air vent, these tubes have a large external diameter and a small bore infusion channel (no larger than that of a single lumen 12 Fr feeding tube).

View Article and Find Full Text PDF

Aims: Hyperlacrimation due to ocular surface or lid abnormality was ruled out and apposition of lower punctum to globe was checked.

Subjects And Methods: This was a pilot study at a tertiary eye care center where 40 patients of proximal canalicular block <8 mm on probing were divided into two equal groups. The external CDCR group underwent routine dacryocystorhinostomy with partial carunculectomy.

View Article and Find Full Text PDF

Small-bore feeding tube versus large-bore sump tube for early enteral nutrition in mechanically ventilated patients: A nationwide inpatient database study.

Clin Nutr

June 2021

Department of Clinical Epidemiology and Health Economics, School of Public Health, The University of Tokyo, 7-3-1 Hongo, Bunkyo-ku, Tokyo 1130033, Japan. Electronic address:

Background & Aims: When physicians start nasogastric tube feeding in mechanically ventilated patients, they have two choices of feeding tube device: a large-bore sump tube or a small-bore feeding tube. Some physicians may prefer to initiate enteral nutrition via the large-bore sump tube that is already in place, and others may prefer to use the small-bore feeding tube. However, it remains unknown whether small-bore feeding tubes or large-bore sump tubes are better for early enteral nutrition.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?