Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Objective: To investigate the effects of 2 components (scintillator and x-ray generator) in the imaging chain on the modulation transfer function (MTF) of a charge-coupled device (CCD) digital intraoral radiographic system.
Study Design: Three screens composed of 3 different scintillator materials, namely europium-doped lutetium oxide transparent optical ceramic (TOC), thallium-doped cesium iodide (CsI), and terbium-doped gadolinium oxysulfide (GOS), were compared. Each was used, in turn, in conjunction with a CCD detector having a pixel dimension of 19.5 mum. Two different x-ray generators were also used to evaluate this variable. MTF was investigated using the slanted slit method.
Results: The TOC provided a good modulation response for low and middle frequencies, reducing to 0 only at a high cutoff frequency. With CsI and GOS, the system MTF dropped to 0 at a lower cutoff frequency than was the case with TOC. Hence, TOC provided higher spatial resolution than the other 2 scintillators tested under the experimental conditions applied. The differences in MTF attributed to the scintillator type were proportional and consistent.
Conclusions: Despite constant pixel dimensions, MTF was affected to a considerable degree by the scintillator applied and the x-ray generator used in conjunction with the same CCD imaging device. TOC shows potential as a possible replacement for CsI and GOS as a scintillator screen material for intraoral digital x-ray imaging using a solid-state detector, providing higher spatial resolution under the given experimental conditions.