Rationale And Objectives: This study addresses the theoretical, experimental, and clinical application of using a central venous catheter system (CVS) for the rapid injection of contrast media during computed tomography (CT).

Methods: Application of Poiseuille's law and the Reynolds equation yielded theoretical data. In-line pressures were measured in experimental models and patients undergoing CT. Diatrizoate meglumine and iohexal contrast media were evaluated.

Results: The Reynolds number was consistent with laminar flow, allowing the application of Poiseuille's law. The calculated and experimental catheter tubing-chamber connection pressures were safe for both contrast media, at rates of 1 mL/second for long catheter tubing and 2 mL/second for short tubing. Thirteen patients had measured pressures within safety limits with no complications.

Conclusions: This study establishes that power injection of contrast media via a CVS during CT is a safe procedure. Benefits include no need for peripheral intravenous catheter placement, decreased paramedical staff radiation exposure, and improved CT image quality.

Download full-text PDF

Source
http://dx.doi.org/10.1097/00004424-199205000-00001DOI Listing

Publication Analysis

Top Keywords

contrast media
20
injection contrast
12
power injection
8
central venous
8
application poiseuille's
8
poiseuille's law
8
contrast
5
media
5
safety considerations
4
considerations power
4

Similar Publications

Role of Cardiovascular MR Imaging and MR Angiography in Patients with Pulmonary Vascular Disease.

Radiol Clin North Am

March 2025

Radiology Department, Northwestern University Feinberg School of Medicine, Arkes Pavilion, 676 North St Clair Street, Suite 800, Chicago, IL 60611, USA. Electronic address:

Cardiac MR imaging and pulmonary MR angiography (MRA) are important clinical tools for the assessment of pulmonary vascular diseases. There are evolving noncontrast and contrast-enhanced techniques to evaluate pulmonary vasculature. Pulmonary MRA is a feasible imaging alternative to CTA in pulmonary embolism detection.

View Article and Find Full Text PDF

Background: Ochratoxin A (OTA) is toxic secondary metabolites produced by fungi and can pose a serious threat to food safety and human health. Due to the high stability and toxicity, OTA contamination in agricultural products is of great concern. Therefore, the development of a highly sensitive and reliable OTA detection method is crucial to ensure food safety.

View Article and Find Full Text PDF

Sentinel lymph node (SLN) detection has been widely investigated in recent years as a part of the surgical staging of women with endometrial cancer (EC), gradually overtaking lymphadenectomy (LND) in this respect. In this study, thirty EC patients, assumed as stage I, were investigated using superparamagnetic iron oxide (SPIO) as a tracer for SLN detection followed by LND. The endpoints of this research were the proportion of successful SLN detection, the average number of SLNs per patient, the percentage of bilaterally detected SLNs, and the proportion of metastatic SLNs.

View Article and Find Full Text PDF

Risk of acute kidney injury following repeated contrast exposure in trauma patients.

Eur J Trauma Emerg Surg

January 2025

Division of Trauma and Emergency Surgery, Department of Surgery, Chang Gung Memorial Hospital, No. 5, Fuxing St., Guishan Dist., Taoyuan City, 333, Taiwan.

Purpose: This study investigates the risk of contrast-associated acute kidney injury (CA-AKI) in trauma patients, focusing on the impact of cumulative contrast medium doses.

Methods: A retrospective review was conducted at a level 1 trauma center (2019-2021). The study included patients who underwent intravascular contrast-enhanced examinations for torso trauma within 7 days post-injury.

View Article and Find Full Text PDF

Technical and clinical reasons for negative skin test results following intradermal testing of contrast media.

Eur J Radiol

January 2025

Department of Diagnostic, Interventional and Pediatric Radiology, Inselspital, University of Bern, Freiburgstrasse, Bern 3010 Switzerland. Electronic address:

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!