AI Article Synopsis

  • The FREDDY laser generates a plasma bubble that produces mechanical shock waves to fragment stones, contrasting with the holmium laser that relies on vaporization for stone destruction.
  • An experimental setup was created to compare the stone retropulsion and fragmentation effects of FREDDY and holmium lasers against a pneumatic lithotripter, using standardized stone phantoms.
  • Both laser types were tested under similar conditions to assess their effectiveness, measuring net retropulsion and fragmentation outcomes after controlled trials with varying energy settings.

Article Abstract

Purpose: The frequency doubled, double pulse Nd:YAG (FREDDY) laser (World of Medicine, Berlin, Germany) functions through the generation of a plasma bubble. Upon bubble collapse a mechanical shock wave is generated, causing stone fragmentation. This mechanism of action is in contrast to the holmium laser, which cause stone destruction by vaporization. Observed clinical stone retropulsion and fragmentation with the FREDDY and holmium lasers has prompted a series of in vitro experiments designed to compare laser induced retropulsion and fragmentation with those of a holmium laser and pneumatic lithotrite.

Materials And Methods: For retropulsion a hands-off underwater laboratory setup, including a horizontally oriented silicone tube 1.3 cm in diameter and a holder to keep the stone phantom in contact with the quartz laser fiber or pneumatic probe, was used. Previously weighed, cylindrical Bego stone phantoms (Bego USA, Smithfield, Rhode Island) were placed in the apparatus. Stone fragmentation was performed with the FREDDY or holmium laser, or the pneumatic lithotripter. The FREDDY and holmium lasers were tested at similar pulse energy and frequency settings. As a standard for comparison, a pneumatic lithotrite was tested with a semirigid probe and single pulse settings of 100, 200 and 300 kPa. Stone phantoms underwent 30 shocks per setting. Mean net retropulsion, defined as the final resting point of the stone, as determined by direct measurement, was recorded for each setting. For fragmentation plaster of Paris stone phantoms of known weights were used to compare the fragmentation ability of each laser. Stones phantoms were placed in a hands-off underwater setup, consisting of an inverted silicon syringe and holder immersed in tap water. The laser fiber (365 microm for the holmium and 280 microm for the FREDDY) was placed through the tip of the syringe in contact with the stone phantom. A total of 24 stones were divided into 4 groups of 6 per group. Two groups were fragmented with the FREDDY laser at 300 and 400 J total energy. The other 2 groups were fragmented using the holmium laser at 300 and 480 J total energy. Fragmentation efficiency was determined as percent weight loss.

Results: For retropulsion at 160 mJ the FREDDY laser caused stone retropulsion to a mean distance of 7.6, 8.1 and 6.8 cm at settings of 5, 10 and 15 Hz, respectively. At 0.8 J the holmium laser retropulsed the stone to a mean distance of 3.3 and 4.9 cm at settings of 5 and 10 Hz, respectively. The pneumatic device caused stone retropulsion a mean distance of 8.5, 9.9 and 13.8 cm at pressure settings of 100, 200 and 300 kPa, respectively. The FREDDY laser generally caused less retropulsion than the pneumatic device, although this difference was only significant at the highest pneumatic lithoclast setting (p <0.05). At clinically relevant settings the FREDDY laser caused significantly more retropulsion than the holmium laser (p <0.05). For fragmentation at total energy settings of 300 and 400 J the FREDDY laser resulted in 44.9% and 86.8% weight loss, respectively (p <0.05). At settings of 300 and 480 J the holmium:YAG laser resulted in 3.3% and 7.1% weight loss, respectively (p <0.05).

Conclusions: At lower frequency settings stone retropulsion was significantly greater with the FREDDY laser compared with the holmium laser. However, retropulsion was significantly less than that caused by the pneumatic lithotripter at all settings. Therefore, we recommend the use of an occlusive device, such as the Stone Cone (Boston Scientific, Natick, Massachusetts) proximal to the calculus during intracorporeal ureteral lithotripsy and in the ureteropelvic junction during percutaneous laser nephrostolithotomy. In vitro stone fragmentation was significantly greater with the FREDDY laser than with the holmium:YAG laser, suggesting that the FREDDY may offer a low cost alternative to the holmium:YAG laser lithotrite in select patients.

Download full-text PDF

Source
http://dx.doi.org/10.1097/01.ju.0000154341.08206.69DOI Listing

Publication Analysis

Top Keywords

holmium laser
20
stone retropulsion
16
freddy laser
16
laser
15
stone
14
retropulsion fragmentation
12
freddy holmium
12
stone phantoms
12
retropulsion
9
fragmentation
8

Similar Publications

The aim of this study was to compare the effectiveness of different types of low level laser treatment (LLLT) in reducing pain levels, changing oxygen saturation and bite force in patients with myofacial pain syndrome (MPS). 45 patients were randomly assigned to three groups: Group 1 (GRR laser, n = 15) received LLLT with Gallium-Aluminium-Arsenide (GaAlAs) diode laser with a wavelength of 904 nm and red laser with a wavelength of 650 nm over masseter muscle region. Group 2 (Nd: YAG laser, n = 15) were treated with Neodymium-doped Yttrium Aluminium Garnet laser with a wavelength of 1064 nm and the same protocol with Nd: YAG laser was performed in the Group 3 (placebo, n = 15) using sham device.

View Article and Find Full Text PDF

Background: We compared the safety and efficacy of holmium laser enucleation of the prostate (HoLEP) and transurethral resection of the prostate (TURP) in elderly men (aged ≥75 years) with benign prostatic hyperplasia (BPH).

Methods: A retrospective analysis of 151 patients (HoLEP: 72; TURP: 79) was conducted. Preoperative and postoperative parameters, including prostate size, International Prostate Symptom Score (IPSS), catheterization duration, hospital stay, and perioperative complications (incontinence and dysuria), were analyzed.

View Article and Find Full Text PDF

Introduction: microRNAs (miRNAs) are small noncoding RNAs and promising cancer biomarkers. Prostate-specific antigen (PSA) testing revolutionized prostate cancer (PCa) diagnostics and monitoring. However, PSA testing also contributes to PCa overdiagnoses that are detrimental on patients' health and may lead to overtreatment.

View Article and Find Full Text PDF

Objectives: To compare postoperative complication rates of patients with metabolic syndrome (MetS) with patients without MetS after holmium laser enucleation of the prostate (HoLEP) for management of benign prostatic hyperplasia (BPH).

Patients And Methods: We retrospectively reviewed patients aged >40 years who underwent HoLEP at our institution from 2007 to 2022. Criteria for MetS were diagnoses of at least three of the following: diabetes mellitus, hypertension, hyperlipidaemia, or obesity (body mass index ≥30 kg/m).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!