Unlabelled: High photon attenuation and scatter in obese patients affect image quality. The purpose of the current study was to optimize lutetium orthosilicate (LSO) PET image acquisition protocols in patients weighing > or =91 kg (200 lb).

Methods: Twenty-five consecutive patients (16 male and 9 female) weighing > or =91 kg (200 lb; range, 91-168 kg [200-370 lb]) were studied with LSO PET/CT. After intravenous injection of 7.77 MBq (0.21 mCi) of 18F-FDG per kilogram of body weight, PET emission scans were acquired for 7 min/bed position. Single-minute frames were extracted from the 7 min/bed position scans to reconstruct 1-7 min/bed position scans for each patient. Three reviewers independently analyzed all 7 reconstructed whole-body images of each patient. A consensus reading followed in cases of disagreement. Thus, 175 whole-body scans (7 per patient) were analyzed for number of hypermetabolic lesions. A region-of-interest approach was used to obtain a quantitative estimate of image quality.

Results: Fifty-nine hypermetabolic lesions identified on 7 min/bed position scans served as the reference standard. Interobserver concordance increased from 64% for 1 min/bed position scans to 70% for 3 min/bed position scans and 78% for 4 min/bed position scans. Concordance rates did not change for longer imaging durations. Region-of-interest analysis revealed that image noise decreased from 21% for 1 min/bed position scans to 14%, 13%, and 11% for, respectively, 4, 5, and 7 min/bed position scans. When compared with the reference standard, 14 lesions (24%) were missed on 1 min/bed position scans but only 2 (3%) on 4 min/bed position scans. Five minute/bed position scans were sufficient to detect all lesions identified on the 7 min/bed position scans.

Conclusion: Lesion detectability and reader concordance peaked for 5 min/bed position scans, with no further diagnostic gain achieved by lengthening the duration of PET emission scanning. Thus, 5 min/bed position scans are sufficient for optimal lesion detection with LSO PET/CT in obese patients.

Download full-text PDF

Source

Publication Analysis

Top Keywords

min/bed position
56
position scans
52
scans
15
position
15
min/bed
14
obese patients
12
lutetium orthosilicate
8
weighing =91
8
=91 200
8
lso pet/ct
8

Similar Publications

Objectives: We investigated image quality and standardized uptake values (SUVs) for different lesion sizes using clinical data generated by F-FDG-prone breast silicon photomultiplier (SiPM)-based positron emission tomography/computed tomography (PET/CT).

Methods: We evaluated the effect of point-spread function (PSF) modeling and Gaussian filtering (Gau) and determined the optimal reconstruction conditions. We compared the signal-to-noise ratio (SNR), contrast, %coefficient of variation (%CV), SUV, and Likert scale score between ordered-subset expectation maximization (OSEM) time-of-flight (TOF) and OSEM+TOF+PSF in phantom and clinical studies.

View Article and Find Full Text PDF

Deep learning-based low count whole-body positron emission tomography denoising incorporating computed tomography priors.

Quant Imaging Med Surg

December 2024

Biomedical Imaging Laboratory (BIG), Department of Electrical and Computer Engineering, Faculty of Science and Technology, University of Macau, Macau, China.

Background: Deep-learning-based denoising improves image quality and quantification accuracy for low count (LC) positron emission tomography (PET). Conventional deep-learning-based denoising methods only require single LC PET image input. This study aims to propose a deep-learning-based LC PET denoising method incorporating computed tomography (CT) priors to further reduce the dose level.

View Article and Find Full Text PDF

Background: Data-driven gated (DDG) PET has gained clinical acceptance and has been shown to match or outperform external-device gated (EDG) PET. However, in most clinical applications, DDG PET is matched with helical CT acquired in free breathing (FB) at a random respiratory phase, leaving registration, and optimal attenuation correction (AC) to chance. Furthermore, DDG PET requires additional scan time to reduce image noise as it only preserves 35%-50% of the PET data at or near the end-expiratory phase of the breathing cycle.

View Article and Find Full Text PDF
Article Synopsis
  • Fibroblast activation protein (FAPI) is gaining attention as a target for imaging various cancers using radiolabeled inhibitors, but no clear imaging protocols have been established yet.
  • This study aimed to determine if the time required for FAPI PET/CT imaging could be shortened from 4 minutes to 3 minutes per bed position without compromising image quality or diagnostic capability.
  • Results showed that while images taken at 3 minutes had acceptable quality, those taken at 1 and 2 minutes were significantly worse than those at 4 minutes, indicating that a reduction in acquisition time could be feasible but with limits to maintain image integrity.
View Article and Find Full Text PDF

Objective: To evaluate the diagnostic utility of 68 Ga-Pentixafor PET/CT for in vivo imaging of CXCR4 receptors in soft tissue/bone sarcoma.

Methods: Ten (7M: 3F; mean age = 24.7 ± 14.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!