Purpose: To determine the possible impact of conductive keratoplasty (CK) on intraocular pressure (IOP) measurements.
Methods: A prospective, single-center, noncomparative interventional case series was performed. Baseline and postoperative IOPs were measured by Goldmann applanation tonometry in 32 eyes of 18 patients who underwent CK for hyperopia correction. Mean follow-up was 11.9 months (range: 8 to 18 months).
Results: After CK, a statistically significant decrease in the measured IOP was observed (before CK: 14.22+/-1.64 vs after CK: 12.66+/-2.21, P<.001). The change in IOP readings postoperatively was not correlated with age, sex, keratometric readings, or attempted correction.
Conclusions: Despite the limitations due to the small number of patients enrolled in this study, the applanation tonometer appears to underestimate the true IOP after CK.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.3928/1081-597X-20050301-13 | DOI Listing |