Patient-specific intensity-modulated radiotherapy (IMRT) verifications require an accurate two-dimensional dosimeter that is not labor-intensive. We assessed the precision and reproducibility of film calibrations over time, measured the elemental composition of the film, measured the intermittency effect, and measured the dosimetric accuracy and reproducibility of calibrated Kodak EDR2 film for single-beam verifications in a solid water phantom and for full-plan verifications in a Rexolite phantom. Repeated measurements of the film sensitometric curve in a single experiment yielded overall uncertainties in dose of 2.1% local and 0.8% relative to 300 cGy. 547 film calibrations over an 18-month period, exposed to a range of doses from 0 to a maximum of 240 MU or 360 MU and using 6 MV or 18 MV energies, had optical density (OD) standard deviations that were 7%-15% of their average values. This indicates that daily film calibrations are essential when EDR2 film is used to obtain absolute dose results. An elemental analysis of EDR2 film revealed that it contains 60% as much silver and 20% as much bromine as Kodak XV2 film. EDR2 film also has an unusual 1.69:1 silver:halide molar ratio, compared with the XV2 film's 1.02:1 ratio, which may affect its chemical reactions. To test EDR2's intermittency effect, the OD generated by a single 300 MU exposure was compared to the ODs generated by exposing the film 1 MU, 2 MU, and 4 MU at a time to a total of 300 MU. An ion chamber recorded the relative dose of all intermittency measurements to account for machine output variations. Using small MU bursts to expose the film resulted in delivery times of 4 to 14 minutes and lowered the film's OD by approximately 2% for both 6 and 18 MV beams. This effect may result in EDR2 film underestimating absolute doses for patient verifications that require long delivery times. After using a calibration to convert EDR2 film's OD to dose values, film measurements agreed within 2% relative difference and 2 mm criteria to ion chamber measurements for both sliding window and step-and-shoot fluence map verifications. Calibrated film results agreed with ion chamber measurements to within 5 % /2 mm criteria for transverse-plane full-plan verifications, but were consistently low. When properly calibrated, EDR2 film can be an adequate two-dimensional dosimeter for IMRT verifications, although it may underestimate doses in regions with long exposure times.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.1852791DOI Listing

Publication Analysis

Top Keywords

edr2 film
28
film
17
imrt verifications
12
film calibrations
12
ion chamber
12
dosimetric accuracy
8
edr2
8
kodak edr2
8
verifications
8
verifications require
8

Similar Publications

Planar dose calculation of electron therapy.

Biomed Phys Eng Express

May 2024

South Carolina Oncology Associates, Columbia, SC 29210, United States of America.

The aim of this study is to determine the planar dose distribution of irregularly-shaped electron beams at their maximum dose depth () using the modied lateral build-up ratio (LBR) and curve-fitting methods.Circular and irregular cutouts were created using Cerrobend alloy for a 14 × 14 cmapplicator. Percentage depth dose (PDD) at the standard source-surface-distance (SSD = 100 cm) and point dose at different SSD were measured for each cutout.

View Article and Find Full Text PDF

Purpose: The purpose of this work is to study the feasibility of using an XRV-124 scintillation detector in measuring the collinearity of the x-ray system and uniform scanning proton beam.

Methods: A brass aperture for Snout 10 was manufactured. The center of the aperture had an opening of 1 cm in diameter (4 cm for the film measurements).

View Article and Find Full Text PDF

Purpose: To investigate the dose-sensitometric response of extended dose range (EDR2) films to scanning carbon-ion beams and to evaluate the applications of the obtained response curves to carbon-ion dose distributions.

Methods: EDR2 films were irradiated by mono-energetic scanning carbon-ion beams with different doses to obtain sensitometric curves at different integrated depth doses (DDDs). Six different DDDs were generated by using a proper buildup for each mono-energetic beam and were used to investigate the energy dependence.

View Article and Find Full Text PDF

Monte Carlo Dose Calculation - A QA Method for SRT and SBRT Plans in Treating Multiple and Small Metastatic Lesions.

J Med Phys

March 2022

Department of Radiation Oncology, Fox Chase Cancer Center, Temple University, Philadelphia, Pennsylvania, USA.

To provide accurate and fast 3-D dose verification for hypofractionated stereotactic radiotherapy (SRT/SBRT) of small and multi targets calculated with a Varian Eclipse treatment planning system (TPS) delivered on a Varian accelerator. Ten brain and lung hypofractionated SRT/SBRT linac-based and CyberKnife plans were generated by the Eclipse system for delivery on the accelerator with the Millenium-120 leaf multileaf collimator (MLC) and Multiplan for the CyberKnife machine. These clinical SRT/SBRT plans required accurate quality assurance measurements to obtain absolute point dose and 3-D dose distributions due to the low number of fractions and high fraction doses.

View Article and Find Full Text PDF

Purpose: To investigate dosimetric implications of biodegradable Biozorb (BZ) markers for proton accelerated partial breast irradiation (APBI) plans.

Materials And Methods: Six different BZs were placed within in-house breast phantoms to acquire computed tomography (CT) images. A contour correction method with proper mass density overriding for BZ titanium clip and surrounding tissue was applied to minimize inaccuracies found in the CT images in the RayStation planning system.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!