Microbiological assay on microtitre plates of folate in serum and red cells.

J Clin Pathol

Department of Haematology, St James's Hospital, Dublin, Ireland.

Published: April 1992

Aims: To develop a simple microbiological assay for serum and red cell folates on 96-well microtitre plates, suitable for use in routine clinical diagnosis.

Methods: Use of a chloramphenicol resistant organism (NCIB 10463) saved time by avoiding aseptic precautions. Use of plate sealers facilitated mixing. Evaluation of assay performance included estimations of folate recovery, assay reproducibility, and response to reduced folate. Results obtained on sera (193) and red cell folates (150) were compared with those obtained using a traditional microbiological assay.

Results: Good recovery of folic acid added to serum and also good interassay and intra-assay precision were obtained with both serum (CV% of less than 5) and red cell folate pools (CV% of less than 5). Equimolar assay responses were obtained with folic acid, 5-formyltetrahydrofolate (L-form), and 5-methyltetrahydrofolate (L-form). The microassay correlated well with a traditional assay for estimation of folate in both serum (n = 193, r = 0.975) and red cells (n = 150, r = 0.96).

Conclusions: This assay is more compact and less time consuming than the traditional assay. It is extremely economical and is easy to perform in a routine clinical laboratory.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC495277PMC
http://dx.doi.org/10.1136/jcp.45.4.344DOI Listing

Publication Analysis

Top Keywords

red cell
12
microbiological assay
8
microtitre plates
8
folate serum
8
serum red
8
red cells
8
cell folates
8
routine clinical
8
folic acid
8
traditional assay
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!