Introduction: Dosimetric studies for targeted radiotherapy require the quantification of activity from scintigraphic images. Quantitative imaging is difficult to achieve because of several effects that can lead to errors in activity estimates, some of which are more apparent when I-131 is considered as a source. An evaluation of these phenomena was performed by modeling the gamma camera and its behavior using Monte Carlo simulations. Two gamma cameras were modeled: DST-XLi and Millennium VG Hawk-Eye (GEMS), and two Monte Carlo codes were used: MCNP (LANL) and GATE (openGate collaboration). GATE is a dedicated single photon emission computed tomography/positron emission tomography (SPECT)/(PET) software based on Geant4 (CERN, Geneve).
Materials And Methods: Gamma-camera modeling was performed in 2 steps: first without a collimator, then with a high-energy, all-purpose (HEAP) collimator according to the specifications given by the manufacturer (the simulation took the hexagonal shape of collimator holes into account). Simulated and measured energy spectra from point sources in air were compared (with or without a collimator). Spatial resolution was obtained from line sources in air at various distances from the detector heads. The photons detected in the 20% energy window from a point source were analyzed in order to determine the amount of primary photons, scattered photons (in the collimator), and septal photons (i.e., photons that crossed the collimator septa without interacting).
Results: Both codes agree well with experimental measurements for the two gamma cameras considered in this study. This allowed us to validate gamma-camera modeling and also served as a benchmark of GATE (new code) versus MCNP (reference code). As shown previously by Dewaraja et al., septal penetration is an important source of image degradation when HEAP collimators are used for I-131 imaging. With the DST-XLi, and for a point source in air, our simulations have shown that 53% of scattered (30%) and septal penetration (23%) photons are detected in the 20% window.
Conclusion: The modeling of two gamma cameras (DST-XLi and Millennium VG Hawk-Eye) has been performed with two Monte Carlo codes (MCNP and Gate). Results obtained with the two Monte Carlo codes agree well with experimental results. As already indicated by several authors, septal penetration and scattered photons in the collimator have a major impact on I-131 scintigraphic imaging.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1089/cbr.2005.20.77 | DOI Listing |
J Chem Theory Comput
January 2025
Department of Chemistry, Rice University, Houston, Texas 77005-1892, United States.
Generalized Hartree-Fock (GHF) is a long-established electronic structure method that can lower the energy (compared to spin-restricted variants) by breaking physical wave function symmetries, namely and . After an exposition of GHF theory, we assess the use of GHF trial wave functions in phaseless auxiliary field quantum Monte Carlo (ph-AFQMC-G) calculations of strongly correlated molecular systems including symmetrically stretched hydrogen rings, carbon dioxide, and dioxygen. Imaginary time propagation is able to restore symmetry and yields energies of comparable or better accuracy than CCSD(T) with unrestricted HF and GHF references, and consistently smooth dissociation curves─a remarkable result given the relative scalability of ph-AFQMC-G to larger system sizes.
View Article and Find Full Text PDFBiomed Opt Express
January 2025
School of Optics and Photonics, Beijing Institute of Technology, Beijing 100081, China.
Radiation therapy (RT) is widely used for cancer treatment but is found with side effects of radiation dermatitis and fibrosis thereby calling for timely assessment. Nevertheless, current clinical assessment methods are found to be subjective, prone to bias, and accompanied by variability. There is, therefore, an unmet clinical need to explore a new assessment technique, ideally portable and affordable, making it accessible to less developed regions too.
View Article and Find Full Text PDFBiomed Opt Express
January 2025
Department of Electronic Engineering, Shanghai Jiao Tong University, Shanghai 200240, China.
Optical coherence tomography angiography (OCTA) offers unparalleled capabilities for non-invasive detection of vessels. However, the lack of accurate models for light-tissue interaction in OCTA jeopardizes the development of the techniques to further extract quantitative information from the measurements. In this manuscript, we propose a Monte Carlo (MC)-based simulation method to precisely describe the signal formation of OCTA based on the fundamental theory of light-tissue interactions.
View Article and Find Full Text PDFBrief Bioinform
November 2024
Department of Biostatistics, University of North Carolina at Chapel Hill, Chapel Hill, NC 27599, United States.
Many human diseases result from a complex interplay of behavioral, clinical, and molecular factors. Integrating low-dimensional behavioral and clinical features with high-dimensional molecular profiles can significantly improve disease outcome prediction and diagnosis. However, while some biomarkers are crucial, many lack informative value.
View Article and Find Full Text PDFOptical polarization is three-dimensional (3-D). Its complete information is described by the nine-component generalized Stokes vector (GSV). However, existing Stokes polarimetry and its design theory are primarily based on the paraxial four-component Stokes vector and 4 × 4 Mueller matrices.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!