Background: The Acorn CorCap cardiac support device (CSD) is a mesh-like device intended to provide end-diastolic support and reduce ventricular wall stress. Animal studies with the CorCap CSD have demonstrated beneficial reverse remodeling, and preliminary safety studies in patients with heart failure have shown that the device is safe and associated with improved left ventricular (LV) structure and function. The objective of the current study was to further evaluate the safety and efficacy of the CorCap CSD in patients with advanced heart failure.
Methods: Twenty-four patients with dilated cardiomyopathy, severe LV dysfunction, and advanced heart failure (NYHA class II-IV) were enrolled at four centers in Italy and Sweden. All patients underwent CorCap CSD implantation either alone (n = 3) or in combination with mitral valve repair/replacement (n = 13), coronary artery bypass surgery (n = 6), combined mitral valve repair/coronary artery bypass surgery (n = 1) or aneurysmectomy (n = 1).
Results: The LV end-diastolic diameter decreased from 69.3 +/- 7.2 to 60.1 +/- 9.0 mm at 3 months, 60.9 +/- 9.6 mm at 6 months, and 58.9 +/- 8.0 mm at 12 months (all p < 0.05). A trend toward an improved LV ejection fraction (28.8 +/- 10.5% at baseline and 32.4 +/- 12.7, 33.1 +/- 10.8, and 33.8 +/- 13.9% at 3, 6 and 12 months respectively) was also noted. The NYHA functional class, 6-min walking distance, and quality of life as measured using Uniscale were all improved. There were no differences in response among the patients submitted to the different types of concomitant surgery.
Conclusions: In agreement with earlier safety studies, even the present investigation demonstrated improvements in cardiac structure and function as well as in patient functional status after Cor Cap CSD implantation. Randomized controlled trials are in progress.
Download full-text PDF |
Source |
---|
Biomed Pharmacother
March 2018
Department of Clinical Pharmacy, School of Basic Medicine and Clinical Pharmacy China Pharmaceutical University, School of Pharmacy, Jiangsu Province, Nanjing 211198, PR China; Department of Heart Surgery, Nanjing Shuiximen Hospital, Jiangsu Province, Nanjing 210017, PR China; Department of Cardiothoracic Surgery, Zhongda Hospital affiliated to Southeast University, Jiangsu Province, Nanjing 210017, PR China. Electronic address:
Complicated pathophysiological syndrome associated with irregular functioning of the heart leading to insufficient blood supply to the organs is linked to congestive heart failure (CHF) which is the leading cause of death in developed countries. Numerous factors can add to heart failure (HF) pathogenesis, including myocardial infarction (MI), genetic factors, coronary artery disease (CAD), ischemia or hypertension. Presently, most of the therapies against CHF cause modest symptom relief but incapable of giving significant recovery for long-term survival outcomes.
View Article and Find Full Text PDFAnn Thorac Surg
June 2013
Department of Mechanical Engineering, University of Kentucky, Lexington, Kentucky, USA.
Background: The Acorn CorCap Cardiac Support Device (CSD; Acorn Cardiovascular Inc, St. Paul, MN) is a woven polyester jacket that is placed around the heart and designed to reverse the progressive remodeling associated with dilated cardiomyopathy. However, the effects of the Acorn CSD on myofiber stress and ventricular function remain unknown.
View Article and Find Full Text PDFAnn Thorac Surg
January 2013
Department of Surgery, University of California at San Francisco and San Francisco Veterans Affairs Medical Center, San Francisco, California, USA.
Background: Myocardial function deteriorates during ventricular remodeling in patients with congestive heart failure (HF). Ventricular restraint therapy using a cardiac support device (CSD) is designed to reduce the amount of stress inside the dilated ventricles, which in turn halts remodeling. However, as an open mesh surrounding the heart, it is unknown what the mechanical properties of the CSD are in different fiber orientations.
View Article and Find Full Text PDFCurr Cardiol Rep
March 2010
Department of Surgery, University of Michigan, Ann Arbor, 48109, USA.
Progressive left ventricular remodeling is characterized by dilation of the left ventricle with a change in the heart from an elliptical shape to a more spherical shape. This change is part of the unfavorable natural history in patients with heart failure. The CorCap Cardiac Support Device (Acorn Cardiovascular, St.
View Article and Find Full Text PDFJ Card Surg
June 2011
Section of Cardiothoracic Surgery & Anesthesiology, Institution of Molecular Medicine and Surgery, Karolinska Institutet, Stockholm, Sweden.
Background And Aim: To evaluate the echocardiographic, functional, and quality of life improving effects of passive containment surgery using the CorCap cardiac support device (CSD; Acorn Cardiovascular Inc., St. Paul, MN, USA) in patients with dilated cardiomyopathy and to investigate the possible differences in ischemic versus idiopathic (i.