Introduction: Alterations in gas displacement in pediatric patients ventilated with the SensorMedics 3100A high-frequency oscillator are most commonly manipulated by adjusting the amplitude, frequency, and percent inspiratory time. The piston-position-and-displacement indicator is commonly centered and subsequently not adjusted. That practice may limit the clinician's ability to optimize carbon dioxide elimination. We hypothesized that varying the piston position would alter gas displacement and carbon dioxide elimination.

Methods: We conducted an observational study in a tertiary pediatric intensive care unit and a correlated bench study. In the clinical study, 24 patients were ventilated with a SensorMedics 3100A high-frequency oscillator. Transcutaneously measured carbon dioxide ((tCO(2))) values were documented with the piston-position-and-displacement indicator in left, center, and right positions. In the bench study the oscillator was set and maintained at: mean airway pressure 15 cm H(2)O, inspiratory time 33% of respiratory-cycle time, bias flow 20 L/min. A pneumotachometer attached to a respiratory mechanics monitor was placed between the ventilator circuit and a test lung. Data were collected with the piston-position-and-displacement indicator at the left, center, and right positions with frequencies of 4-14 Hz and amplitudes of 25-55 cm H(2)O. Data were collected over a 3-minute time period for each combination of frequency, amplitude, and piston-position-and-displacement-indicator position. We compared the data with repeated-measures analysis of variance. Pairwise comparisons were performed with a 2-tailed Student's test with Bonferroni correction.

Results: Among the 24 patients (tCO(2)) was significantly associated with the position of the piston (p < 0.007). In the bench study, gas displacement was higher when the piston-position-and-displacement indicator was positioned to the left (than when at the center position) 91.7% of the time (p < 0.0001). When the piston-position-and-displacement indicator was positioned to the right (as compared to the center position), gas displacement was lower 75% of the time (p < 0.0001).

Conclusion: Adjusting the oscillator piston alters the volume of gas displaced and provides an additional means for titrating carbon dioxide elimination. .

Download full-text PDF

Source

Publication Analysis

Top Keywords

carbon dioxide
20
gas displacement
20
piston-position-and-displacement indicator
20
dioxide elimination
12
bench study
12
left center
12
piston position
8
patients ventilated
8
ventilated sensormedics
8
sensormedics 3100a
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!