Hydrodynamics-induced variability in the USP apparatus II dissolution test.

Int J Pharm

Department of Chemical and Biochemical Engineering, Rutgers University, 98 Brett Road, Piscataway, NJ 08854-8058, USA.

Published: March 2005

The USP tablet dissolution test is an analytical tool used for the verification of drug release processes and formulation selection within the pharmaceutical industry. Given the strong impact of this test, it is surprising that operating conditions and testing devices have been selected empirically. In fact, the flow phenomena in the USP test have received little attention in the past. An examination of the hydrodynamics in the USP apparatus II shows that the device is highly vulnerable to mixing problems that can affect testing performance and consistency. Experimental and computational techniques reveal that the flow field within the device is not uniform, and dissolution results can vary dramatically with the position of the tablet within the vessel. Specifically, computations predict sharp variations in the shear along the bottom of the vessel where the tablet is most likely to settle. Experiments in which the tablet location was carefully controlled reveal that the variation of shear within the testing device can affect the measured dissolution rate.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ijpharm.2004.08.003DOI Listing

Publication Analysis

Top Keywords

usp apparatus
8
dissolution test
8
hydrodynamics-induced variability
4
usp
4
variability usp
4
dissolution
4
apparatus dissolution
4
test
4
test usp
4
tablet
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!