The accuracy of the new SNAP index with the Bispectral index (BIS) to distinguish different states of propofol/remifentanil anaesthesia was compared in 19 female patients who were undergoing minor gynaecological surgery. Comparisons of the SNAP index, BIS, spectral edge frequency, mean arterial blood pressure and heart rate were performed. The ability of all parameters to distinguish between the steps of anaesthesia -awake vs. loss of response, awake vs. anaesthesia, anaesthesia vs. first reaction and anaesthesia vs. extubation - were analysed with the prediction probability. The prediction probability to differentiate between two interesting nuances of anaesthetic states -loss of response vs. first reaction - was calculated. Only the BIS showed no overlap between the investigated steps of anaesthesia. Both the SNAP index and BIS failed to differentiate the nuances of anaesthesia. The SNAP index and BIS were superior to mean arterial blood pressure and heart rate and spectral edge frequency in distinguishing between different steps of anaesthesia with propofol and remifentanil and provided useful additional information.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1365-2044.2004.04120.xDOI Listing

Publication Analysis

Top Keywords

snap bis
12
steps anaesthesia
12
anaesthesia
9
snap bispectral
8
states propofol/remifentanil
8
propofol/remifentanil anaesthesia
8
spectral edge
8
edge frequency
8
arterial blood
8
blood pressure
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!