Objective: To determine the accuracy of 2 commercially available point-of-care devices relative to plasma international normalized ratio (INR) values.

Patients And Methods: Point-of-care INR testing was performed with the CoaguChek and ProTime 3 devices in consecutive patients attending an anticoagulation clinic between June 18, 2003, and August 6, 2003. Results were compared with plasma INRs using a sensitive thromboplastin (International Sensitivity Index, 1.0).

Results: Ninety-four patients agreed to participate in the study. Relative to the plasma INR, values were in agreement +/-0.4 INR unit 82% and 39% of the time for the CoaguChek and ProTime 3 devices, respectively. The mean +/- SD CoaguChek INRs were 0.2+/-0.31 unit lower, whereas ProTime 3 INRs were 0.8+/-0.68 unit higher than plasma INR values. Treatment decisions based on these data would have resulted in inappropriate dose adjustments 10% and 22% of the time for these 2 respective devices. Correlation with plasma was greater for the CoaguChek (r2=0.90) compared with the ProTime 3 device (r2=0.73).

Conclusions: Optimal warfarin treatment requires accurate measurement of the INR. The choice of a point-of-care device for INR management depends on the reliability of INR data generated by the device.

Download full-text PDF

Source
http://dx.doi.org/10.4065/80.2.181DOI Listing

Publication Analysis

Top Keywords

international normalized
8
normalized ratio
8
relative plasma
8
inr
8
coaguchek protime
8
protime devices
8
plasma inr
8
inr values
8
plasma
5
device
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!