[Transesophageal arterial oxygen saturation monitoring: an experimental study].

Sichuan Da Xue Xue Bao Yi Xue Ban

Department of Anesthesiology, West China Hospital, Sichuan University, Chengdu 610041, China.

Published: January 2005

Objective: To assess the sensitivity and accuracy of a novel transesophageal approach to monitoring the descending aortic oximetry (SeO2).

Methods: Nine dogs were involved in the experimental study. After the induction of anaesthesia, the carrier of the oximetry probe (Nellcor-D20, USA) was inserted into the lower segment of esophagus to monitor SeO2, and the probe was "locked" in position of post-descending aorta after the opening of thoracic cavity. Another probe was pasted on the surface of lingual mucous membrane. The readings and figures of SeO2 and surface of lingual mucosa oximetry (SmO2) were observed continuously and recorded simultaneously. Vital signs were monitored with pulse oxygen saturation (SpO2), invasive blood pressure by femoral artery, HR, EKG, PetCO2, T, FiO2. The SaO2 of blood gas analysis by femoral artery was used as the "gold standard" to calculated the relative and absolute deviations of SeO2 and SmO2. The changes of SeO2 and SaO2 were compared in case of acute hypoxia when values of SmO2 dropped to 90%, 80%, 70%, 60% and the patient was re-ventilated with 100% oxygen.

Results: (1) SeO2, SmO2 and SaO2 were 100% when the patients were ventilated with 100% oxygen. During hypoxia, the descent of SeO2 from 100% to 90% was (91.03+/-20.23) s (P<0.001) earlier than that of SmO2. And after re-supply of pure oxygen, the ascent of SeO2 was (25.9+/-6.0) s (P<0.05) earlier than that of SmO2. (2) SaO2 was well related with SeO2 and SmO2 (R2: 0.9884 and 0.9296) respectively. The relative and absolute deviations of SeO2 were 1.6% and 1.3%, while those of SmO2 were 7.6% and 6.1% from arterial blood samples SaO2. (3) There were no significant differences in MAP, HR, ECG, PetCO2 and T.

Conclusion: This study showed that SeO2 monitoring is sensitive. It could accurately reflect the arterial oxygen saturation not only in normal condition but also during hypoxia and the re-ventilation with 100% oxygen. SeO2 responds faster and is closer to SaO2, compared with SmO2 measurements. This may be an alternative method in the cases where the monitoring of peripheral SpO2 is difficult.

Download full-text PDF

Source

Publication Analysis

Top Keywords

oxygen saturation
8
surface lingual
8
femoral artery
8
seo2 smo2
8
seo2
6
[transesophageal arterial
4
arterial oxygen
4
saturation monitoring
4
monitoring experimental
4
experimental study]
4

Similar Publications

The aim of this study was to compare the effectiveness of different types of low level laser treatment (LLLT) in reducing pain levels, changing oxygen saturation and bite force in patients with myofacial pain syndrome (MPS). 45 patients were randomly assigned to three groups: Group 1 (GRR laser, n = 15) received LLLT with Gallium-Aluminium-Arsenide (GaAlAs) diode laser with a wavelength of 904 nm and red laser with a wavelength of 650 nm over masseter muscle region. Group 2 (Nd: YAG laser, n = 15) were treated with Neodymium-doped Yttrium Aluminium Garnet laser with a wavelength of 1064 nm and the same protocol with Nd: YAG laser was performed in the Group 3 (placebo, n = 15) using sham device.

View Article and Find Full Text PDF

Background: The circadian rhythm of myocardial infarction (MI) in patients with obstructive sleep apnea (OSA) remains disputable and no studies have directly evaluated the relationship between nocturnal hypoxemia and the circadian rhythm of MI. The aim of the current study was to evaluate the association of OSA and nocturnal hypoxemia with MI onset during the night.

Methods: Patients with MI in the OSA-acute coronary syndrome (ACS) project (NCT03362385) were recruited.

View Article and Find Full Text PDF

Purpose: To study the effects of breathing exercises on preventing pulmonary complications in patients undergoing coronary artery bypass graft surgery.

Methods: Observing whether preoperative breathing exercises can reduce the incidence of postoperative pulmonary complications in patients undergoing coronary artery bypass graft surgery; observing whether these exercises can improve postoperative arterial oxygen pressure, oxygen saturation, and the distance walked in a six-minute walk test after surgery; as well as reduce hospital stay duration, lower treatment costs, and improve the quality of life as measured by the Short Form-36 Health Survey (SF-36).

Design: The study population includes patients undergoing coronary artery bypass graft surgery under general anesthesia; the research center is Capital Medical University Xuanwu Hospital; the sample size is 120.

View Article and Find Full Text PDF

Transposition of great arteries (TGA) is a critical congenital heart disease leading to a fatal outcome if timely management is not provided. Management in low-income countries is challenging. A retrospective analysis was carried out at Sudan Heart Center for infants with TGA who underwent balloon atrial septostomy (BAS) from January 2010 to December 2020.

View Article and Find Full Text PDF

Pain Perception Threshold in Young High-Altitude Natives After Acute Exposure to Severe Hypoxic Conditions.

Oxygen (Basel)

March 2025

Centro de Investigación en Medicina de Altura (CIMA), Facultad de Medicina Humana, Universidad de San Martín de Porres, Lima 15001, Peru.

Previous studies indicate that individuals at high altitudes have a lower pain threshold than those living at sea level. This study evaluates the differences in pain perception among young people living at an altitude of 3800 m and after acute exposure to a severe hypoxic environment at more than 5100 m. Fourteen people (BMI of 22.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!