A one- or two-dimensional high performance liquid chromatography system for electrospray ionization mass spectrometers has been developed that is optimized for ion exchange and reversed phase separations. A unique and simple valve configuration permits the use of a variety of non-volatile salts; ammonium sulfate was used in an example of strong cation exchange separations. The system was designed and evaluated for both micro- and nanoflow chromatography. The peptide detection limit was approximately 100 fmol for micro- and 20 fmol for nanoflow, demonstrating the concentration and mass sensitivity improvements expected with nanoelectrospray ionization. The 1D/2D-HPLC MS system is fully automated for routine peptide analyses, compatible with direct injection of proteolytic digests, and exhibits chromatographic reproducibility and sensitivity. Software permits operator selection of either a 1D or 2D configuration with corresponding system parameters as required for individual samples. The hardware elements and resulting performance are described in this paper.