Automatic gain control in mass spectrometry using a jet disrupter electrode in an electrodynamic ion funnel.

J Am Soc Mass Spectrom

Biological Sciences Division, Pacific Northwest National Laboratory, Richland, Washington 99352, USA.

Published: February 2005

We report on the use of a jet disrupter electrode in an electrodynamic ion funnel as an electronic valve to regulate the intensity of the ion beam transmitted through the interface of a mass spectrometer in order to perform automatic gain control (AGC). The ion flux is determined by either directly detecting the ion current on the conductance limiting orifice of the ion funnel or using a short mass spectrometry acquisition. Based upon the ion flux intensity, the voltage of the jet disrupter is adjusted to alter the transmission efficiency of the ion funnel to provide a desired ion population to the mass analyzer. Ion beam regulation by an ion funnel is shown to provide control to within a few percent of a targeted ion intensity or abundance. The utility of ion funnel AGC was evaluated using a protein tryptic digest analyzed with liquid chromatography Fourier transform ion cyclotron resonance (LC-FTICR) mass spectrometry. The ion population in the ICR cell was accurately controlled to selected levels, which improved data quality and provided better mass measurement accuracy.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC1989159PMC
http://dx.doi.org/10.1016/j.jasms.2004.11.003DOI Listing

Publication Analysis

Top Keywords

ion funnel
24
ion
15
mass spectrometry
12
jet disrupter
12
automatic gain
8
gain control
8
disrupter electrode
8
electrode electrodynamic
8
electrodynamic ion
8
ion beam
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!