Objectives: We investigated whether embolic particles could be detected as high-intensity transient signals (HITS) with a Doppler guide wire during percutaneous coronary intervention (PCI) in patients with acute myocardial infarction (AMI) We also assessed whether these signals could be reduced using a distal protection (DP) device.
Background: Embolization of thrombi and plaque components to the microcirculation is a major complication of PCI in patients with AMI. Embolic particles running in the cerebral artery are detected as HITS by transcranial Doppler ultrasound.
Methods: We prospectively studied 16 consecutive patients with AMI who underwent direct PCI within 24 h after the onset of symptoms. A PercuSurge GuardWire (MedtronicAVE, Santa Rosa, California) was used as the DP device. Eight patients were randomly assigned to the non-DP group, and the remaining eight were assigned to the DP group. Coronary flow velocity was recorded continuously from before the first balloon inflation to after balloon deflation.
Results: All patients in the non-DP group had HITS detected (12 +/- 9 counts) within five consecutive beats (4 +/- 1 beat) after balloon deflation, but none were detected in any of the patients in the DP group.
Conclusions: The Doppler guide wire can be used to visually detect and count emboli as HITS, and the DP device is effective for prevention of distal embolization.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jacc.2004.09.062 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!