Aim: The objective of this study was to measure the microbial contamination released from dental unit air lines (DUAL) and dental unit water lines (DUWL).

Materials And Methods: Emissions from DUAL and DUWL from five dental units supplied by a central water source (tap) and a centralised air supply were sampled three times over a five- week period. Air was forced through sterile water and then plated onto selective agar using apparatus designed to spread the sample solution evenly, and then incubated at room temperature for seven days. Colonies were then counted and the concentration of bacteria present was determined and expressed as colony forming units per millilitre (cfu/ml) per minute. The same procedure was used to evaluate five other dental units, which had attached independent water reservoir (bottle) systems (IWR). Only deionised water was added to the bottles and unit waterlines were cleaned weekly with alkaline peroxide based solution. Values were expressed as cfu/ml.

Results: Air and water specimens obtained from dental units supplied with tap water had microbial counts significantly (t-test, p < 0.05) greater than IWR dental units.

Conclusions: Results indicate that IWR can reduce the numbers of micro-organisms released from DUWL. However, the effectiveness of such systems should be routinely monitored. Emissions from DUAL seems to reflect the levels of microbial contamination present in DUWL.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1875-595x.2004.tb00301.xDOI Listing

Publication Analysis

Top Keywords

microbial contamination
12
dental unit
12
dental units
12
unit air
8
water
8
air water
8
water lines
8
emissions dual
8
units supplied
8
dental
7

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!