We describe a patient in whom the bispectral index (BIS) decreased to 0 during surgery. A 42-yr-old man with chronic renal failure was scheduled to undergo construction of an arteriovenous shunt. He had a history of acute cerebral hemorrhage. An intracranial hematoma had been removed a month earlier with almost complete neurological recovery. He had uncontrolled hypertension. His systolic blood pressure was 180 mm Hg before anesthesia induction. Anesthesia was induced with 100 mg of propofol and 3% sevoflurane. After laryngeal mask insertion, anesthesia was maintained with nitrous oxide 60% in oxygen and sevoflurane. BIS decreased to near 0 on 2 occasions: after anesthesia induction and shortly after the start of the surgery. His systolic blood pressure decreased to 110 mm Hg and BIS increased when his blood pressure was increased to 130-140 mm Hg. The decrease in BIS was suspected to be the result of decreased cerebral blood flow. The systolic blood pressure of 110 mm Hg (mean blood pressure, 80 mm Hg) was probably less than the lower limit of autoregulation. Although BIS has some limitations as a cerebral monitor, it was useful for detecting possible cerebral hypoperfusion in this case.

Download full-text PDF

Source
http://dx.doi.org/10.1213/01.ANE.0000139347.64944.95DOI Listing

Publication Analysis

Top Keywords

blood pressure
20
systolic blood
12
cerebral hypoperfusion
8
bis decreased
8
anesthesia induction
8
blood
6
anesthesia
5
bis
5
pressure
5
detection cerebral
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!