Background: To evaluate the per-operative process of shoulderjoint replacement, time-action analysis can be used.

Material/methods: Forty procedures performed by 7 surgeons with different experience using 5 different prostheses and 3 different surgical approaches were analyzed.

Results: The surgical procedures showed a large variation in, for example, duration, tasks of team members, and protocol used. The surgical procedure was influenced by several factors, such as the prosthesis used, the surgical approach, the patient's condition, and the experience of the surgeon. Exposure of the glenoid was difficult and several retractors were needed, which were held by an extra assistant or clamped to the table or the surgeon. Two main limitations were seen in all procedures: repeated actions and waiting. Also, five errors could be identified. None of the alignment instruments was completely reliable and they allowed the surgeon to make major errors.

Conclusions: Better alignment instruments, pre-operative planning techniques, and operation protocols are needed for shoulder prostheses. The training of resident surgeons should be focused on the exposure phase, the alignment of the humeral head, the exposure of the glenoid, and the alignment of the glenoid. Evaluating the surgical process using time-action analysis can be used to determine the limitations during surgical procedures. Furthermore, it shows the large variation in factors affecting surgical performance, indicating that a system approach is needed to improve surgical outcome.

Download full-text PDF

Source

Publication Analysis

Top Keywords

time-action analysis
12
surgical
9
surgical process
8
replacement time-action
8
surgical procedures
8
procedures large
8
large variation
8
exposure glenoid
8
alignment instruments
8
factors influencing
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!