A process to fabricate nonplanar microelectrode array circuits was developed and the microelectrodes were characterized. These platinum microelectrode arrays are for recording streaming potential signals generated during indentation of articular cartilage. The nonplanar substrate was produced by permanent deformation of a 7-in-diameter circular stainless-steel wafer to form 32 semi-spherical caps (radius of curvature = 4.65 mm and height = 250 microm) at the periphery. The wafer was covered with a 2.5-microm-thick layer of insulating polyimide. Standard microelectronic processes were applied to produce 32 circuits (60 mm long x 4 mm wide) with 37 exposed circular microelectrodes (diameter = 100 microm) centered over each semi-spherical cap. A 2.5-microm-thick photodefinable polyimide layer encapsulated the conducting lines. Capacitances between one microelectrode and either another microelectrode or the metallic substrate were 14.6 +/- 2.0 and 34.4 +/- 3.3 pF, respectively, at 100 Hz. The impedance of the microelectrodes in a 0.15 M saline bath (PBS) was 0.25 +/- 0.08 Mohms while the crosstalk (Vinduced/Vapplied) between two microelectrodes was 0.20 +/- 0.11%, at 100 Hz. Indentation measurements were performed on articular cartilage in vitro showing,streaming potentials that indicate electrode-tissue contact times and generation of streaming potentials.

Download full-text PDF

Source
http://dx.doi.org/10.1109/TBME.2004.836522DOI Listing

Publication Analysis

Top Keywords

nonplanar microelectrode
8
microelectrode array
8
array circuits
8
articular cartilage
8
microelectrode
5
fabrication characterization
4
characterization nonplanar
4
circuits arthroscopic
4
arthroscopic diagnosis
4
diagnosis cartilage
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!