Computational wear prediction of a total knee replacement from in vivo kinematics.

J Biomech

Department of Mechanical & Aerospace Engineering, University of Florida, Gainesville, FL 32611, USA.

Published: February 2005

AI Article Synopsis

  • The wear of ultra-high molecular weight polyethylene bearings in knee replacements is a significant factor limiting their longevity.
  • The study utilizes computer modeling with fluoroscopic data to predict patient-specific implant damage by simulating different activities and loads.
  • Findings show that the predicted wear patterns closely match actual damage observed in retrieved implants, suggesting this modeling can enhance the design and longevity of knee replacements.

Article Abstract

Wear of ultra-high molecular weight polyethylene bearings in total knee replacements remains a major limitation to the longevity of these clinically successful devices. Few design tools are currently available to predict mild wear in implants based on varying kinematics, loads, and material properties. This paper reports the implementation of a computer modeling approach that uses fluoroscopically measured motions as inputs and predicts patient-specific implant damage using computationally efficient dynamic contact and tribological analyses. Multibody dynamic simulations of two activities (gait and stair) with two loading conditions (70-30 and 50-50 medial-lateral load splits) were generated from fluoroscopic data to predict contact pressure and slip velocity time histories for individual elements on the tibial insert surface. These time histories were used in a computational wear analysis to predict the depth of damage due to wear and creep experienced by each element. Predicted damage areas, volumes, and maximum depths were evaluated against a tibial insert retrieved from the same patient who provided the in vivo motions. Overall, the predicted damage was in close agreement with damage observed on the retrieval. The gait and stair simulations separately predicted the correct location of maximum damage on the lateral side, whereas a combination of gait and stair was required to predict the correct location on the medial side. Predicted maximum damage depths were consistent with the retrieval as well. Total computation time for each damage prediction was less than 30 min. Continuing refinement of this approach will provide a robust tool for accurately predicting clinically relevant wear in total knee replacements.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jbiomech.2004.02.013DOI Listing

Publication Analysis

Top Keywords

total knee
12
gait stair
12
computational wear
8
knee replacements
8
damage
8
time histories
8
tibial insert
8
predicted damage
8
correct location
8
maximum damage
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!