Background: Roche Diagnostics has issued new c-fas calibrators for its automated systems. These produce creatinine values that are more comparable with those obtained by high-performance liquid chromatography. However, this results in an underestimation of measured creatinine at concentrations below 155 micromol/L and an overestimation at concentrations above this value.

Methods: Serum and urine creatinine concentrations were prospectively determined on samples from 60 patients using the new (compensated) and old (uncompensated) c-fas calibrators, and Passing-Bablok regression analysis was performed. The regression equations thus determined were then used retrospectively to determine the compensated creatinine results (i.e. those results that would have been obtained using the new calibrator) in those serum and urine samples analysed in the previous year using the old uncompensated c-fas calibrator. The compensated creatinine results were then used to estimate the glomerular filtration rate (GFR) by calculating creatinine clearance. This was done by using the formula: UV/Pt, in which U represents the urinary creatinine concentration (micromol/L), V the urinary collection volume (mL), P the serum creatinine concentration (micromol/L) and t the urinary collection time (min). It was also calculated using the abbreviated Modification of Diet in Renal Disease study group (MDRD) formula.

Results: The creatinine clearance as determined using either the UV/Pt calculation or the MDRD formula overestimated GFR by approximately 30% and approximately 50%, respectively, in normal individuals with a serum creatinine concentration below 155 micromol/L. However, in patients with mild to moderate renal failure (serum creatinine from 155 to 500 micromol/L), changes in creatinine clearances determined by the two procedures were minimal.

Conclusion: When laboratories introduce this new, compensated calibrator into practice, it may be appropriate to discuss its potential impact with clinical staff who monitor patients using creatinine clearance.

Download full-text PDF

Source
http://dx.doi.org/10.1258/0004563042466776DOI Listing

Publication Analysis

Top Keywords

creatinine
14
creatinine clearance
12
creatinine concentration
12
serum creatinine
12
glomerular filtration
8
filtration rate
8
c-fas calibrators
8
creatinine concentrations
8
155 micromol/l
8
serum urine
8

Similar Publications

Background: Serum and urinary uromodulin are emerging as potential cardiovascular risk factors. The aim of our study was to determine uromodulin in both serum and urine to evaluate their potential as early cardiovascular risk markers and markers of kidney function in children and young adults.

Methods: This case-control study included 72 participants - 42 children and young adults with chronic kidney disease stages 1-2 and 30 healthy controls.

View Article and Find Full Text PDF

The case report presents a male patient in his mid-60s with a history of hypertension, benign prostatic hyperplasia and chronic kidney disease (CKD). He presented with gradually increasing serum creatinine levels and hyperglobulinemia, leading to suspicion of multiple myeloma. However, subsequent testing revealed features consistent with systemic lupus erythematosus (SLE) and IgG4-related kidney disease (IgG4-RKD).

View Article and Find Full Text PDF

Objective: This study aimed to investigate the relationship between mortality and the frontal QRS-T angle (FQRS-TA), obtained by calculating the absolute difference between the QRS and T waves electrocardiographically (ECG), in patients diagnosed with ischemic stroke (IS).

Methods: This research is a retrospective and cross-sectional study. The diagnosis of IS was confirmed through brain imaging and physical examination.

View Article and Find Full Text PDF

Hepatic hydrothorax (HH) is a severe cirrhosis complication requiring early diagnosis and appropriate management. This study aimed to assess the impact of HH on the disease severity and mortality of cirrhotic patients and compare their clinical and biological profiles with those of patients without HH. This retrospective study involved 155 patients diagnosed with cirrhosis, of whom 31 had HH.

View Article and Find Full Text PDF

Renal disease is common in patients with cardiovascular disease (CVD) and is associated with adverse outcomes. Cardiac magnetic resonance (CMR) with advanced mapping techniques is the gold standard for characterizing myocardial tissue, and renal tissue is often visualized on these maps. However, it remains unclear whether renal T1 times accurately reflect renal dysfunction or predict adverse outcomes.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!