Clinical photoablation with a 500-Hz scanning spot excimer laser.

J Refract Surg

Institut für Refraktive und Okuloplastische Chirurgie (IROC AG), Zollikerstrasse 164, CH-8008 Zurich, Switzerland.

Published: January 2005

Purpose: The aim of this study was to use a 500-Hz scanning spot laser (Concept500, WaveLight Laser Technologie AG, Erlangen, Germany) to investigate potential side effects that might be associated with the use of a high repetition rate laser platform.

Methods: Seven eyes were treated using a 500-Hz scanning spot laser for laser in situ keratomileusis (LASIK). The local frequency of the ablation was kept below 40 Hz to avoid local heating of corneal tissue. With the exception of the high repetition rate (500 Hz), all other laser parameters such as fluence, algorithm, ablation profile, and spot diameter were identical to a standard WaveLight Allegretto laser system. Patients were examined at 1 month and 1 year after initial treatment. Preoperative and postoperative examination included manifest sphere and cylinder, uncorrected and best spectacle-corrected visual acuity (BSCVA).

Results: All eyes were treated for myopia or myopic astigmatism. Five eyes received spherocylindrical and two eyes spherical ablation only. No adverse events correlated with the use of a high repetition rate laser system were observed during surgery or at any point during follow-up. All eyes maintained or had improved BSCVA at 12 months after treatment when compared to preoperative values.

Conclusion: The use of an excimer laser with a maximal repetition rate of 500 Hz and a local repetition rate of less than 40 Hz was free of any specific side effect that might be associated with the use of such a high repetition rate.

Download full-text PDF

Source
http://dx.doi.org/10.3928/1081-597X-20041101-12DOI Listing

Publication Analysis

Top Keywords

repetition rate
24
high repetition
16
500-hz scanning
12
scanning spot
12
laser
10
excimer laser
8
spot laser
8
associated high
8
rate laser
8
eyes treated
8

Similar Publications

Recent advances in Light Emitting Diode (LED) technology have enabled a more affordable high frame rate photoacoustic imaging (PA) alternative to traditional laser-based PA systems that are costly and have slow pulse repetition rate. However, a major disadvantage with LEDs is the low energy outputs that do not produce high signal-to-noise ratio (SNR) PA images. There have been recent advancements in integrating deep learning methodologies aimed to address the challenge of improving SNR in LED-PA images, yet comprehensive evaluations across varied datasets and architectures are lacking.

View Article and Find Full Text PDF

In fluvial environments, the shifting of river channels and bank erosion are frequently caused by both natural and anthropogenic factors. Riverine hazards like bank erosion and course alterations offer severe issues to the riparian villages along the lower basin of the Tista River in India, which substantially influence the livelihoods of inhabitants living there. This research addressed river channel shifting tendency and identified major bank erosion-prone villages along the lower course of the Tista River and challenges to the livelihoods of the riparian people.

View Article and Find Full Text PDF

Supercontinuum generation in scintillator crystals.

Sci Rep

January 2025

Laser Research Center, Vilnius University, Saulėtekio Avenue 10, LT-10223, Vilnius, Lithuania.

We present a comparative experimental study of supercontinuum generation in undoped scintillator crystals: bismuth germanate (BGO), yttrium orthosilicate (YSO), lutetium oxyorthosilicate (LSO), lutetium yttrium oxyorthosilicate (LYSO) and gadolinium gallium garnet (GGG), pumped by 180 fs fundamental harmonic pulses of an amplified Yb:KGW laser. In addition to these materials, experiments in yttrium aluminium garnet (YAG), potassium gadolinium tungstate (KGW) and lithium tantalate (LT) were performed under identical experimental settings (focusing geometry and sample thickness), which served for straightforward comparison of supercontinuum generation performances. The threshold and optimal (that produces optimized red-shifted spectral extent) pump pulse energies for supercontinuum generation were evaluated from detailed measurements of spectral broadening dynamics.

View Article and Find Full Text PDF

To address the challenge of reusing foaming agents in foam drainage gas production processes, we developed a redox-responsive surfactant with a straightforward preparation method based on molecular electrostatic interaction assembly. The redox response mechanism of the surfactant was investigated through surface tension, absorbance, particle size, and Zeta potential analyses. Results indicate that the minimum surface tension in the oxidized state can reach 26.

View Article and Find Full Text PDF

Compact wearable microstrip antenna design using hybrid quasi-Newton and Taguchi optimization.

Sci Rep

January 2025

Department of Computer Science and Engineering, Symbiosis Institute of Technology, Symbiosis University Pune, Pune, India.

A novel approach is introduced for designing a miniaturized wearable antenna. Utilizing Taguchi's philosophy typically entails numerous experimentations runs, but our method significantly reduces these by employing a quasi-Newton approach with gradient descent to estimate process parameter ranges. This hybrid technique expedites convergence by streamlining experiments.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!