An HPLC-APCI-MS(/MS) method for the (trace) analysis of the most commonly encountered peroxide explosives, hexamethylenetriperoxidediamine (HMTD) and triacetonetriperoxide (TATP), has been developed. With this method, HMTD and TATP have been analyzed in the same run. (Pseudo-)molecular ions of these peroxides have been obtained as base peak under the same condition. A series of product ions was produced from these pseudo-molecular ions ([HMTD - 1]+ and [TATP + NH4]+) in the MS/MS analysis. We also pioneered in showing that a TATP molecular ion [TATP + H]+ can be observed with HPLC-MS/MS. The limit of detection for HMTD and TATP was 0.26 and 3.3 ng, respectively, on column by HPLC-MS in the Full Scan mode and 0.08 and 0.8, respectively, by HPLC-APCI-MS/MS in Selected Reaction Monitoring (single mass unit) mode. The method presented has been applied successfully for the identification of peroxides in the bulk solid state (powder sample), as well as in post-blast extracts originating from a forensic case. For the post-blast extracts, the use of tandem MS has been shown clearly to be of crucial importance for the identification and detection of the peroxide explosives.