Objective: The usefulness of skin-level gastrostomy tubes (button systems) for maintaining nutrition support after primary insertion of percutaneous endoscopic gastrostomy (PEG) is widely accepted. After first promising experiences with newly developed skin-level gastrostomy system (Freka-Button Gastrostomy) the safety and long-term stability of this new tube was not defined.

Material And Methods: In a cohort study we prospectively evaluated from 2.1998 until 12.2001 for ease of use, complications, time to failure and long-term follow-up of 61 Freka button gastrostomies inserted in 50 patients (mean age 57.6 years, range 6 - 78 years, 44 men, 6 women). Mean time after primary PEG placement was 6.3 months (range 1 - 30 months).

Results: Correct application of all buttons (48 gastric, 2 jejunal) was easy and fast (median time 11 minutes) to perform. In 56 % we used a new Seldinger guide wire technique to improve stomal passage. Within the first ten days and during long-term follow-up (median 248, range 2 - 593 days) no major complication was seen. 11 systems had to be replaced mainly due to balloon failure (median 352, range 186 - 593 days). The total observation time were 15,128 days with a system failure rate of 0.26 per year.

Conclusion: The Freka-button systems provides an easy-to-use, safe, and feasible alternative for long-term use of skin-level gastrostomy. First jejunal application of this new device was promising but needs further evaluation.

Download full-text PDF

Source
http://dx.doi.org/10.1055/s-2004-813782DOI Listing

Publication Analysis

Top Keywords

skin-level gastrostomy
12
gastric jejunal
8
long-term follow-up
8
593 days
8
long-term
5
gastrostomy
5
[skin-level gastrostomy
4
gastrostomy long-term
4
long-term prospective
4
prospective trial
4

Similar Publications

In this Festschrift communication, the author of two novel approaches to long-term enteral access: the gastrostomy without laparotomy (Percutaneous Endoscopic Gastrostomy or PEG) and the skin-level feeding device (Gastrostomy "Button"), relates the history of these innovations, describes the initial and extended applications, and comments on the implications of these developments.

View Article and Find Full Text PDF

Laparoscopic-assisted percutaneous endoscopic gastrostomy: insertion of a skin-level device using a tear-away sheath.

Can J Surg

August 2015

From the Division of General Surgery, Schulich School of Medicine & Dentistry, Western University (Livingston, Pepe, Jones, Bütter, Merritt); and the Division of Paediatric Surgery, Schulich School of Medicine & Dentistry, Western University (Jones, Bütter, Merritt), London, Ont.

Background: This study describes our experience with the placement of a skin-level gastrostomy device (MIC-KEY) in a single procedure.

Methods: We identified infants, children and young adults who underwent laparoscopic-assisted percutaneous endoscopic gastrostomy (LAPEG) tube insertion between October 2009 and June 2013. The steps of this procedure include upper endoscopy, single-port laparoscopy, gastropexy via percutaneous T-fasteners and placement of a skin-level gastrostomy device (MIC-KEY) using a "push" technique with a tear-away sheath.

View Article and Find Full Text PDF

Ancillary procedures necessary for translational research in experimental craniomaxillofacial surgery.

J Craniofac Surg

November 2014

*Department of Plastic & Reconstructive Surgery, Johns Hopkins University School of Medicine; Departments of †Otolaryngology-Head & Neck Surgery and ‡Radiology, 3D Medical Applications Center, Naval Postgraduate Dental School, Walter Reed National Military Medical Center, Bethesda; §Robotics and Autonomy Group, Research and Engineering Development Department, The Johns Hopkins University Applied Physics Laboratory, Laurel; ∥Laboratory for Computational Sensing & Robotics, Johns Hopkins University; and Departments of ¶Molecular and Comparative Pathobiology and #Surgery, Johns Hopkins University School of Medicine, Baltimore, MD.

Introduction: Swine are often regarded as having analogous facial skeletons to humans and therefore serve as an ideal animal model for translational investigation. However, there is a dearth of literature describing the pertinent ancillary procedures required for craniomaxillofacial research. With this in mind, our objective was to evaluate all necessary procedures required for perioperative management and animal safety related to experimental craniomaxillofacial surgical procedures such as orthotopic, maxillofacial transplantation.

View Article and Find Full Text PDF

Background/purpose: Various techniques have been presented to remove the percutaneous endoscopically placed gastrostomy tube in children, but tubes with semi-rigid internal retaining discs are difficult or impossible to remove by external traction. We describe a simple and effective endoscopic removal technique that should be applicable to any type of percutaneous endoscopic gastrostomy tube.

Methods: Percutaneous endoscopic gastrostomy tube removal was performed with the "tie and retrograde pull" technique.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!