Nonpulsatile blood pumps are mainly used in extracorporeal life support systems. Although pulsating blood flow is known to be physiological, a pulsatile pump is not commonly applied in a circuit with a membrane oxygenator because of damage to the blood cells. The hypothesis that the placement of a pressure-relieving compliance chamber in a circuit might reduce blood cell trauma was tested. An extracorporeal life support circuit was constructed in an acute lung injury model of dogs by oleic acid infusion. The animals were divided into three groups. In group I (n = 6) a nonpulsatile centrifugal pump was used as a control. In group II (n = 4) a single-pulsatile pump was used, and in group III (n = 6) a single-pulsatile pump equipped with a compliance chamber was used. Pump flow was maintained at 1.8-2.0 L/min for 2 h. Hemodynamics and blood gas analyses indicated that the pulsatile groups II and III had better results than the nonpulsatile group I. The plasma-free hemoglobin level, which indicates blood cell trauma, was the lowest in group I and the highest in group II but was significantly decreased in group III. A pressure-relieving compliance chamber could significantly reduce high circuit pressures and blood cell trauma.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1525-1594.2004.07341.xDOI Listing

Publication Analysis

Top Keywords

compliance chamber
16
extracorporeal life
12
life support
12
blood cell
12
cell trauma
12
pressure-relieving compliance
8
single-pulsatile pump
8
group iii
8
blood
7
group
7

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!