Hypothesis: Ossiculoplasty using prosthetic reconstruction with a malleus assembly to the stapes head will result in better transmission of vibrations from the eardrum to the stapes footplate than reconstruction with a tympanic membrane assembly to the stapes head. Both types of reconstruction will be affected by tension of the prosthesis.

Background: Theories (and some clinical studies) that the shape of the normal tympanic membrane is important suggest that prosthetic reconstruction to the malleus performs better than reconstruction to the tympanic membrane. This has not been previously tested by directly measuring vibration responses in the human ear. Our previous work suggests that tympanic membrane assembly to the stapes head type prostheses performed best under low tension. This had not been previously tested for malleus assembly to the stapes head type prostheses.

Methods: Hydroxyapatite prostheses were used to reconstruct a missing incus defect in a fresh cadaveric human ear model. Two types of prostheses were used, one from the stapes head to the malleus (malleus assembly to the stapes head), the other from the stapes head to the tympanic membrane (tympanic membrane assembly to the stapes head). Stapes footplate center responses were measured using a laser Doppler vibrometer in response to calibrated acoustic frequency sweeps.

Results: Tension had a very significant effect on both types of prostheses in the lower frequencies. Loose tension was best overall. The malleus assembly to the stapes head type prostheses consistently performed better than the tympanic membrane assembly to the stapes head type prostheses when stratified for tension.

Conclusion: Tension has a significant effect on prosthesis function. Malleus assembly to the stapes head type prostheses generally result in better transmission of vibrations to the stapes footplate than tympanic membrane assembly to the stapes head type prostheses.

Download full-text PDF

Source
http://dx.doi.org/10.1097/00129492-200411000-00008DOI Listing

Publication Analysis

Top Keywords

stapes head
48
assembly stapes
40
tympanic membrane
32
head type
24
malleus assembly
20
membrane assembly
20
type prostheses
20
stapes
15
head
14
stapes footplate
12

Similar Publications

To investigate the occurrence and managements of poor recovery after total endoscopic middle ear surgery. A total of 302 cases(315 ears) who underwent endoscopic middle ear surgery in our hospital from June 2020 to June 2021 were collected. Follow up by means of endoscopy, pure tone hearing threshold, tympanogram was conducted at 1 month, 3 months, 6 months and 1 year after surgery to analyze the incidence, possible causes, treatment strategies and effects of poor results tympanic membrane healing and hearing recovery.

View Article and Find Full Text PDF

Objective: To determine which ear environment risk factors impact ossiculoplasty hearing outcomes and to generate a statistically-valid grading system for ossiculoplasty outcome reporting.

Study Type: Retrospective case series.

Methods: A multi-institutional database was generated from cases performed between 2011 and 2019.

View Article and Find Full Text PDF

[Comparison of microdrill versus diode laser in endoscopic stapedotomy].

Zhonghua Yi Xue Za Zhi

December 2024

Department of Otolaryngology-Head and Neck Surgery, Eye and ENT Hospital of Fudan University, Shanghai 200031, China.

To compare the surgical time, auditory outcomes and safety between microdrill and diode laser in endoscopic stapedotomy. The data of patients with otosclerosis who underwent endoscopic stapedotomy in Eye and ENT Hospital of Fudan University from January 2020 to December 2023 were retrospectively analyzed. The patients were divided into microdrill group and laser group according to treatment method.

View Article and Find Full Text PDF

Background: The "floating mass transducer" (FMT) of the Vibrant Soundbridge was initially attached to the long process of the incus (LP). In patients missing an incus, coupling to the stapes head using the Vibroplasty CliP-coupler is used. Its main limitation is the height, leading to potential contact with the tympanic membrane.

View Article and Find Full Text PDF

Objective: To explore the historical evolution of surgical techniques for otosclerosis treatment, viewed through the lens of human temporal bone pathology to aid in understanding the disease and refining surgical interventions.

Data Sources: A review of historical literature on otosclerosis, surgical techniques, and otopathological findings was conducted. Eight temporal bone specimens from the Paparella Otopathology & Pathogenesis Laboratory, University of Minnesota, and one from the University of California, Los Angeles, were analyzed.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!