The Bryan Cervical Disc: wear properties and early clinical results.

Spine J

Department of Orthopedic Surgery and Rehabilitation, University of Wisconsin Hospitals, 600 Highland Avenue, K4/735, Madison, WI 53792, USA.

Published: February 2005

Background Context: The rationale for motion preservation by disc replacement after anterior cervical discectomy is to diminish long-term morbidity secondary to adjacent segment degeneration. However, these disc prostheses will be subject to wear and its possible adverse consequences.

Purpose: Assess the in vitro and in vivo wear properties of the Bryan Cervical Disc and the early clinical results.

Study Design: In vitro mechanical testing, caprine animal model and prospective European human trial.

Outcome Measure: In vitro mechanical testing used American Society for Testing and Materials standards to measure wear rates and debris. Caprine results were analyzed by histopathologic analysis by a blinded veterinary pathologist. Clinical outcomes were measured by Odoms's criteria, independent radiologic analysis and Short Form-36.

Methods: In vitro wear testing in cervical spine simulator for up to 40 M cycles. Wear rates were determined and particles analyzed microscopically. The in vivo inflammatory response was studied in goats that had discectomy at C4-C5. They were sacrificed between 3 and 12 months and histopathologically were compared with controls who had fusions with titanium plates. A prospective clinical trial of 136 patients with minimum 1-year follow-up was analyzed using validated questionnaires and radiographs.

Results: In vitro wear averaged approximately 1.76% by weight at 10 M cycles and 18% at 40 M cycles. Wear debris were present in the periprosthetic tissues in 4 of 11 animals without inflammatory response. The early clinical results were satisfactory in over 90% of patients.

Conclusion: The in vivo and in vitro wear properties are satisfactory for the expected duration of life of the prosthesis. The early clinical results are satisfactory and equal to fusion.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.spinee.2004.07.026DOI Listing

Publication Analysis

Top Keywords

early clinical
16
wear properties
12
vitro wear
12
wear
9
bryan cervical
8
cervical disc
8
vitro mechanical
8
mechanical testing
8
wear rates
8
cycles wear
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!