Background: Sources of cyanide exposure are many, including combustion of plastic and vinyl, such as in a house fire, laboratory or industrial exposures including exposure in the electroplating industry both of printed circuit boards and in jewelry work. Rapid and definitive diagnosis of cyanide poisoning is unavailable in the emergency department setting. It is desirable to make a definitive diagnosis in order to prevent potential complications of empiric treatment of presumptive cyanide poisoning from the cyanide antidote kit currently approved by the US Food and Drug Administration (FDA). We investigated a technique to detect cyanide currently utilized by water treatment facilities to determine if it can be applied to rapidly detect concentrations of cyanide in the clinically important range.

Methods: Varying standardized dilutions of KCN ranging from 0.25 microg/mL to 30 microg/mL were acidified with a drop of sulphuric acid in a closed system under a ventilation hood. Cyantesmo test strips were placed into the test tubes above the fluid level where liberated HCN gas interacted with the test strip to effect a color change. Color changes were compared to negative controls and to each other.

Results: The test strips demonstrated an incrementally increasing deep blue color change over a progressively longer portion of the test strip in less than 5 minutes for each concentration of KCN including 1, 3, 10, and 30 microg/mL. The concentrations of 0.25, 0.5, and 0.75 required more than 2 hours to begin demonstration of any color change.

Conclusion: The Cyantesmo test strips accurately and rapidly detected, in a semi-quantifiable manner, concentrations of CN greater than 1 microg/mL contained in each test sample. Future work to validate this test in blood and in clinical specimens is planned.

Download full-text PDF

Source
http://dx.doi.org/10.1081/clt-200035349DOI Listing

Publication Analysis

Top Keywords

test strips
12
definitive diagnosis
8
cyanide poisoning
8
test
8
cyantesmo test
8
test strip
8
color change
8
cyanide
6
rapid cyanide
4
cyanide detection
4

Similar Publications

A novel Schirmer Strip-Based Tear Matrix Metalloproteinase Measurement in Dry Eye Evaluation.

Ocul Surf

January 2025

Department of Ophthalmology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences; Key Laboratory of Ocular Fundus Diseases, Chinese Academy of Medical Sciences. Electronic address:

Unlabelled: Dry eye disease is a multifactorial disorder of the ocular surface with increasing global prevalence, yet no universally accepted "gold standard" exists for its diagnosis or severity assessment. Tear matrix metalloproteinase 9 (MMP-9) is widely recognized as a valuable biomarker for dry eye, yet there remains a critical need for a simple, accurate, and broadly applicable method for its quantification. This study aims to develop and evaluate a Schirmer strip-based Eu-time resolved fluorescence immunochromatography (Eu-TRFICO) method for the quantitative detection of MMP-9 in tears.

View Article and Find Full Text PDF

Purpose: To compare the color alteration, surface roughness and microhardness and cross-sectional microhardness of bovine enamel treated with at-home whitening strips and gels.

Materials And Methods: Sixty-six pigmented specimens (n = 11) were allocated to six groups: C-cotton wool moistened with distilled water for 1 h; SDS-sodium dithionite strip, for 1 h; HPS-6.5% hydrogen peroxide strip, for 1 h; CPS-20% carbamide peroxide strip, for 1 h; HPG-7.

View Article and Find Full Text PDF

Lab-on-paper for molecular testing with USB-powered isothermal amplification and fluidic control.

Mikrochim Acta

January 2025

Department of Biotechnology and Bioengineering, Chonnam National University, Gwangju, 61186, Republic of Korea.

The global healthcare market increasingly demands affordable molecular diagnostics for field testing. To address this need, we introduce a lab-on-paper (LOP) platform that integrates isothermal amplification with a specially designed paper strip for molecular testing through an automated microfluidics process. The LOP system is engineered for rapid, cost-effective, and highly sensitive detection, using USB-powered thermal management and a wax valve mechanism.

View Article and Find Full Text PDF

Current American Urological Association guidelines recommend that patients with acute obstructive kidney stone requiring continuous anticoagulation/antiplatelet therapy should not be treated by shockwave lithotripsy or percutaneous nephrolithotomy because of the risk of catastrophic renal hemorrhage possible with those techniques. Currently, ureteroscopy is the only recommended surgical treatment. We evaluated if burst wave lithotripsy (BWL) could be used in these cases by treating pigs with BWL while undergoing anticoagulation therapy.

View Article and Find Full Text PDF

Reliability of various Antimicrobial Susceptibility Testing Methods for Piperacillin/tazobactam in challenging Escherichia coli isolates.

J Glob Antimicrob Resist

January 2025

Department of Medical Microbiology and Infectious Diseases, Erasmus University Medical Center, Rotterdam, The Netherlands. Electronic address:

Piperacillin/tazobactam antimicrobial susceptibility testing (AST) against Enterobacterales can be challenging. The aim of this study was to assess the reproducibility of various automated (Vitek®2) and non-automated AST methods (broth microdilution (BMD), minimum inhibitory concentration (MIC) test strip, and disk diffusion) for piperacillin/tazobactam in 'challenging' E. coli isolates.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!